Calibrator Output Operations
2
Keysight U1401B User’s and Service Guide
47
Memory Generation
For constant voltage and current outputs, the U1401B offers two additional useful
functions. One is an
autoscan
output that is able to generate up to 16 different
steps of constant voltage or current each with its own user-defined amplitude and
time interval. The other one is an
autoramp
output with user-defined dual slopes
and number of steps for linear simulation.
Autoscan output
To set the
autoscan
output:
1
Turn the rotary switch to any one of the
(constant current output) or
(constant voltage output) positions.
2
Press
SHIFT
to access the shifted operations of the keypad. The
annunciator will appear on the display.
3
Follow one of the instructions below:
– For voltage output, press
MODE
to cycle through ±1.5 V, ±15 V,
±1.5 V,
±15 V,
±1.5 V, and
±15 V output modes.
Select one of the two
output modes, depending on the voltage
range you require.
– For current output, press
MODE
to cycle through ±25 mA,
±25 mA,
and
±25 mA output modes. Select the
output mode.
Содержание U1401B
Страница 1: ...Keysight U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 10: ...10 Keysight U1401B User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 61: ...Making Measurements 3 Keysight U1401B User s and Service Guide 61 Figure 3 1 DC voltage measurement ...
Страница 76: ...3 Making Measurements 76 Keysight U1401B User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 107: ...Application Examples 5 Keysight U1401B User s and Service Guide 107 Figure 5 5 Zener diode test ...
Страница 128: ...5 Application Examples 128 Keysight U1401B User s and Service Guide Figure 5 17 Verifying a frequency transmitter ...