Calibrator Output Operations
44
Keysight U1401B User’s and Service Guide
Enabling and Disabling the Output
The U1401B can generate and measure signals simultaneously. Pressing the
OUTPUT key disables the U1401B output by placing it in the standby mode.
Pressing this key again toggles the output on.
When the output is in the standby mode, the
annunciator disappears and the
annunciator is displayed instead. This means the calibrator has stopped
generating its output.
The standby mode will also be activated automatically if:
– you accidentally feed an external signal into the output terminals while the
output function is enabled.
– the noise from an external power system or output terminals causes an error
signal to the output. For example, when ESD is performed with a voltage of
8000 V, this instrument will go into standby mode.
– an overload condition has been detected when generating constant voltage or
square wave outputs.
– the weak or low battery condition occurs. This ensures output quality and
serves as another alert to let the user know that the energy level of the
batteries is low.
– you put the slide switch into the
M
(input only) position (you should do this to
conserve battery power if you do not intend to use any of the output
functions).
Содержание U1401B
Страница 1: ...Keysight U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 10: ...10 Keysight U1401B User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 61: ...Making Measurements 3 Keysight U1401B User s and Service Guide 61 Figure 3 1 DC voltage measurement ...
Страница 76: ...3 Making Measurements 76 Keysight U1401B User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 107: ...Application Examples 5 Keysight U1401B User s and Service Guide 107 Figure 5 5 Zener diode test ...
Страница 128: ...5 Application Examples 128 Keysight U1401B User s and Service Guide Figure 5 17 Verifying a frequency transmitter ...