Calibration Procedures
1
Keysight U1241C/U1242C Service Guide
51
8
When all calibrations have completed, press both
and
together for more than 1 second to exit the calibration mode. Alternatively,
you can also power off and on the multimeter again. The multimeter will return
to the normal measurement mode and is secured.
9
Verify the adjustments using the
“Performance Verification Tests”
Exiting the adjustment mode
1
Remove all the shorting plugs and connectors from the multimeter.
2
Record the new calibration count.
3
Press both
and
together for more than 1 second
to exit the
calibration mode.
4
Cycle the multimeter’s power. The multimeter will then be secured.