Performance Tests and Calibration
5
Keysight U1241AN User’s and Service Guide
59
Recommended Test Equipment
The test equipment recommended for the performance verification and
adjustment procedures is listed below. If the exact equipment is not available,
substitute the calibration standards of equivalent accuracy.
Table 5-1
Recommended test equipment
Application
Recommended Equipment
Recommended Accuracy Requirements
DC Voltage
Fluke 5520A
< 1/5 instrument 1 year specifications
DC Current
Fluke 5520A
< 1/5 instrument 1 year specifications
AC Voltage
Fluke 5520A
< 1/5 instrument 1 year specifications
AC Current
Fluke 5520A
< 1/5 instrument 1 year specifications
Resistance
Fluke 5520A
< 1/5 instrument 1 year specifications
Frequency
Fluke 5520A
< 1/5 instrument 1 year specifications
Capacitance
Fluke 5520A
< 1/5 instrument 1 year specifications
Diode
Fluke 5520A
< 1/5 instrument 1 year specifications
Temperature
Fluke 5520A
< 1/5 instrument 1 year specifications
TM Electronics KMPC1MP
(K-Type thermocouple extension)
—
Short
Pomona MDP-S
< 1/5 instrument 1 year specifications
Содержание U1241AN
Страница 1: ...Keysight U1241AN Handheld Digital Multimeter User s and Service Guide ...
Страница 14: ...14 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 28: ...1 Getting Started 28 Keysight U1241AN User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...