Getting Started
1
Keysight U1211A, U1212A, and U1213A User’s and Service Guide
31
The terminals at a glance
Figure 1-7
Clamp meter terminal inputs
WARNING
Ensure that the terminal connections are correct for a particular
measurement before making any measurement. To avoid damage to the
device, do not exceed the input limit.
Red input
terminals (
V
, ,
, )
Black input
terminal (
COM
)
Table 1-4
Terminal connections for different measurement functions
Measurement functions
Input terminals
Input limit
AC current
Clamp jaw
1000 A
rms
DC current
[a]
AC voltage
V
COM
CAT III 1000 V
rms
CAT IV 600 V
rms
DC voltage
Resistance
COM
1000 V
rms
for
short circuit < 0.3 A
Capacitance
Diode
Temperature
[b]
[a] DC current measurement is only available for U1212A and U1213A clamp meters.
[b] Temperature function is only available for U1212A and U1213A clamp meters.