Keysight N5244A/45A Service Guide 4-43
Troubleshooting
Measurement System Troubleshooting
5.
If the 800 MHz signal is not present and the analyzer does not have
mechanical switches, replace the reference coupler. Refer to
and Replacing the A29–A32 Reference Couplers and Reference Coupler
Mounting Brackets” on page 7-42
6.
If the 800 MHz signal is present and the analyzer has source attenuators
and bias tees, reconnect the cable to the reference coupler and continue
testing at
“Checking the A38, A39, A40, and A41 60-dB Source Step
7.
If the 800 MHz signal is present and the analyzer does not have source
attenuators and bias tees, replace the test port coupler. Refer to
“Removing and Replacing the A33–A36 Test Port Couplers” on page 7-44
Checking the A38, A39, A40, and A41 60-dB Source Step Attenuators
1.
Locate the appropriate semirigid cable at the output of the source step
attenuator to be checked:
—
Options 219, 224, 419, and 423
—
Port 1; W82 of A38
—
Port 2; W94 of A41
—
Port 3; W86 of A39
—
Port 4; W90 of A36
2.
Using a 5/16-inch torque wrench, disconnect the semirigid cable at the
step attenuator.
3.
Connect the spectrum analyzer to the open step attenuator connector. Set
the spectrum analyzer to measure a signal at 800 MHz.
4.
If the 800 MHz signal is not present, replace the source step attenuator.
“Removing and Replacing the A38–A41 Source Attenuators and
the A46–A49 Receiver Attenuators” on page 7-48
.
5.
If the 800 MHz signal is present, replace the associated bias tee. Refer to
“Removing and Replacing the A38–A41 Source Attenuators and the
A46–A49 Receiver Attenuators” on page 7-48
Checking the A50, A51, A52, and A53 Mechanical Switches
1.
Locate the appropriate semirigid cable at the output of the mechanical
switch to be checked:
—
Options 224 and 423
—
Port 1; W106 of A50
—
Port 2; W120 of A53
—
Port 3; W112 of A51
—
Port 4; W116 of A52
Содержание PNA-X N5244A
Страница 10: ...8 Keysight N5244A 45A Service Guide Contents ...
Страница 90: ...3 62 Keysight N5244A 45A Service Guide Tests and Adjustments Adjustments 3 ...
Страница 146: ...4 56 Keysight N5244A 45A Service Guide Troubleshooting Instrument Block Diagrams Sheet 4 4 ...
Страница 155: ...Keysight N5244A 45A Service Guide 5 9 Theory of Operation Synthesized Source Group Operation Figure 5 3 Source Group ...
Страница 174: ...5 28 Keysight N5244A 45A Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 6 4 Port Receiver Group ...
Страница 175: ...Keysight N5244A 45A Service Guide 5 29 Theory of Operation Receiver Group Operation Figure 5 7 2 Port Receiver Group ...
Страница 188: ...5 42 Keysight N5244A 45A Service Guide Theory of Operation Noise Measurement Group Operation Option 029 5 ...
Страница 247: ...Keysight N5244A 45A Service Guide 6 59 Replaceable Parts Replaceable Parts Listings This page intentionally left blank ...
Страница 311: ...Keysight N5244A 45A Service Guide 6 123 Replaceable Parts Replaceable Parts Listings ...
Страница 313: ...Keysight N5244A 45A Service Guide 6 125 Replaceable Parts Replaceable Parts Listings This page intentionally left blank ...
Страница 323: ...Keysight N5244A 45A Service Guide 6 135 Replaceable Parts Replaceable Parts Listings This page intentionally left blank ...
Страница 343: ...Keysight N5244A 45A Service Guide 6 155 Replaceable Parts Replaceable Parts Listings This page intentionally left blank ...
Страница 363: ...Keysight N5244A 45A Service Guide 6 175 Replaceable Parts Replaceable Parts Listings This page intentionally left blank ...
Страница 472: ...8 18 Keysight N5244A 45A Service Guide General Purpose Maintenance Procedures Correction Constants 8 ...
Страница 504: ...32 Keysight N5244A 45A Service Guide Index ...
Страница 505: ...Installation Note Xxxxx xxxxx 3 ...