Keysight P937xA PXIe Service Guide 63
Removal and Replacement Procedures
Connector Replacement Procedure
Connector Replacement Procedure
Parts Required
— SMA or 3.5 mm connectors
— Nuts for connectors
Tools Required
— “Spark plug” adapter
— Torque wrench (for “spark plug” adapter)
— ESD equipment:
Removal Procedure
1. Place the module on its side on an ESD safe, flat surface.
2. Place the “spark plug” adapter on the nut of the connector you are going to
replace. The adapter must fit snugly into the slots of the nut. Refer to
Figure 5-1
Tools and Parts
Description
Keysight Part Number
ESD grounding wrist strap
9300-1367
5-ft grounding cord for wrist strap
9300-0980
2 x 4 ft conductive table mat and 15-ft grounding wire
9300-0797
ESD heel strap (for use with conductive floors)
9300-1308
Although the photos in this section show a 3.5 mm connector being
replaced, the procedure is the same for SMA connectors.