Keysight P937xA PXIe Service Guide 49
Tests, Adjustments, and Troubleshooting
Adjustments
Source Adjustment
The source calibration is used to adjust your network analyzer for a flat source
power across its full frequency range. There are differences between each test
port; therefore, an adjustment is required for each port.
Equipment Used for the Source Adjustment
Procedure
1. Connect the USB power sensor and adapter to the analyzer Port 1. If using
a power meter, make the proper GPIB connections. Refer to
Hardware Required for your GPIB Instruments” on page 27
.
2. Press
Utility > System > Service > Adjustments
.
3. In the Select Desired Adjustment dialog box, click
Source Adjustment
.
4. Follow the instructions and prompts as they are displayed. If using a power
meter, ensure the GPIB settings for the sensor(s) are correct in the dialog
box.
Equipment Type
Model or Part Number
Alternate Model or Part Number
USB Thermocouple power sensor U8485A Option 200
U8485A power meter
Adapter, 3.5 mm -m- to 3.5 mm
-m-
83059A
85052-60014