CAT (Cable and Antenna Test) Mode
CAT Mode Settings
—
Left
The Settings table is shown to the left of the trace window
—
Bottom
The Settings table is shown below the trace window
— When finished changing ALL settings, press
Done
to save your settings
Frequency Range
Set the range of frequencies over which you would like to make CAT Mode
measurements
When the frequency range is changed after a calibration is performed, the cal
becomes interpolated. Learn more in
.
How to set Frequency Range
— Press
Freq/Dist
.
— Then choose from the following
—
Start
and
Stop
frequencies - beginning and end of the sweep.
—
Center
and
Span
frequencies – the center frequency and span of
frequencies (half on either side of center).
— Follow each setting by entering a value using the numeric keypad,
arrows, or the rotary knob.
—After using the keypad, select a multiplier key. Learn about multiplier
abbreviations in
“Multiplier Abbreviations” on page 34
—After using the
arrows or the rotary knob, press
Enter
. The
amount of frequency increment is based on the current span and can
NOT be changed in CAT Mode.
Scale Settings
Adjust the Y-axis scale to see the relevant portions of the data trace. The Y-axis
is divided into 10 graticules.
This setting can be changed at any time without affecting calibration accuracy.
How to set Scale
— Press
Scale / Amptd
.
— Then choose from the following three methods:
1.
Autoscale All
Autoscales all of the traces on the screen, useful only
for dual-trace configurations.
2. Set Scale, Reference Level, and Reference Position:
—
Scale
Manually enter a
scale per division
to view specific
areas of the trace
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...