186 Keysight N9923-90001 User’s Guide
Data Analysis Features
All about Limit Lines
——
Upper
- FAILS any measured data point
ABOVE
the line.
——
Lower
- FAILS any measured data point
BELOW
the line.
—
Off/On
– Press
Edit
to toggle between the following:
——
On
- Limit line is drawn on the screen and PASS/FAIL testing
occurs.
——
Off
- Limit line is NOT drawn on the screen and testing does
NOT occur.
—
Frequency
– Enter a numeric X-axis value of the data point.
—
Amplitude
– Enter a numeric Y-axis value of the data point.
— Press
Add
then choose from the following:
—
Add Point
- Adds a blank X/Y data point below the cursor.
—
Add Limit
- Adds a default limit line below the current limit.
— Press
Clear
then choose from the following:
—
Clear Point
- Removes the X/Y data point on which the cursor resides.
—
Clear Limit
- Removes the limit Line on which the cursor resides.
—
Clear All
- Removes ALL limit lines.
— Press
<Back
to return to the Limit menu.
Relative Limit Lines
Relative limit lines are X/Y coordinates that are referenced to the center
frequency (X-axis) and Reference level (Y-axis).
Use relative limits to test devices or signals that are identical except for the
center frequency or amplitude. With relative limit lines in place, you change the
center frequency of the measurement and the limits look as though they are in
the same location.
The same is true for Reference Level. When you expect to measure a different
power level, instead of changing the Y-axis value of all of the fixed limits,
simply change to relative limits. The limits will then remain a constant Y-axis
distance from the Reference Level.
Each limit line can be changed at any time from Fixed to Relative, and Relative
to Fixed.
How to Create a Relative Limit Line
The easiest way to create a Relative limit line is to create a Fixed Limit at a
specific center frequency and reference level. Learn how
. Then change the limit line to Relative. The X/Y data points will be
re-calculated relative to the center frequency and reference level.
Содержание N9923A
Страница 1: ...Keysight Technologies FieldFox RF Network Analyzers N9923A User s Guide ...
Страница 14: ...14 Keysight N9923 90001 User s Guide Contents ...
Страница 20: ...20 Keysight N9923 90001 User s Guide Overview FieldFox Manuals Software and Supplemental Help ...
Страница 36: ...36 Keysight N9923 90001 User s Guide Preparing for Initial Use of Your New FieldFox Connector Care ...
Страница 60: ...60 Keysight N9923 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 100: ...100 Keysight N9923 90001 User s Guide DTF Distance to Fault Measurements DTF Measurement Settings ...
Страница 124: ...124 Keysight N9923 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 134: ...134 Keysight N9923 90001 User s Guide Power Meter USB Mode USB Power Meter Settings ...
Страница 144: ...144 Keysight N9923 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 160: ...160 Keysight N9923 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 206: ...206 Keysight N9923 90001 User s Guide File Management Printing ...
Страница 246: ...246 Keysight N9923 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 264: ...264 Keysight N9923 90001 User s Guide Specifications Data Sheet ...
Страница 267: ...267 Keysight N9923A Handheld Analyzer User s Guide D Korean Battery Safety Translation ...
Страница 268: ...268 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 269: ...Keysight N9923 90001 User s Guide 269 Korean Battery Safety Translation ...
Страница 270: ...270 Keysight N9923 90001 User s Guide Korean Battery Safety Translation ...
Страница 275: ...Installation Note Xxxxx xxxxx 3 ...