Keysight Technologies N9040B UXA Signal Analyzer Service Guide 459
Assembly Replacement Procedures
Motherboards
10.
. Remove the front bracket
(1)
by removing the
nine screws
(2)
(0515-0372).
Figure 15-106
Front Bracket Removal
11.
Remove the Low Noise Path and Microwave Preselector Bypass Switches.
“Low Noise Path and Microwave Preselector Bypass Switches”
removal procedure.
12.
Remove the Front End Assembly. Refer to the
removal procedure.
13.
. Remove the seven screws
(1)
(0515-0372) that
attach the front motherboard bracket to the chassis. Use the ejectors to
disconnect from the rear motherboard and pull out towards the front of
the instrument to remove from the chassis.
Содержание N9040B Option 513
Страница 4: ...4 ...
Страница 138: ...138 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Messages Condition Messages ...
Страница 240: ...240 Chapter 11 Hardware Options A21 Wideband Analog IF Troubleshooting Figure 11 1 Analog IF Block Diagram ...
Страница 256: ...256 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 306: ...306 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 2 External Hardware ...
Страница 338: ...338 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 28 Chassis ...
Страница 346: ...346 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts ...
Страница 502: ...502 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Post Repair Procedures ...
Страница 532: ...532 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Functional Tests Scale Fidelity ...
Страница 536: ...536 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Software Software Updates ...
Страница 541: ...Index Index 541 Y yellow LED 48 51 yig filter see A12 YTF YTF see A12 YTF ...