Keysight Technologies N9040B UXA Signal Analyzer Service Guide 353
Assembly Replacement Procedures
Instrument Outer Case
Replacement
1.
Disconnect the instrument from ac power.
2.
Slide the instrument cover back onto the deck from the rear. The seam on
the cover should be on the bottom. Be sure the cover seats into the gasket
groove in the Front Frame Assembly.
3.
Replace the four rear feet to the rear of the instrument. Torque the rear
feet screws (0515-1619 and 3050-0893 washers) to 21 inch-pounds.
4.
Replace the bottom feet by sliding into place until they snap in. Install the
locks by pressing in flat.
5.
Replace the handle straps on both sides of the instrument. Torque the
handle strap screws to 21 inch-pounds.
Содержание N9040B Option 513
Страница 4: ...4 ...
Страница 138: ...138 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Messages Condition Messages ...
Страница 240: ...240 Chapter 11 Hardware Options A21 Wideband Analog IF Troubleshooting Figure 11 1 Analog IF Block Diagram ...
Страница 256: ...256 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 306: ...306 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 2 External Hardware ...
Страница 338: ...338 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 28 Chassis ...
Страница 346: ...346 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts ...
Страница 502: ...502 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Post Repair Procedures ...
Страница 532: ...532 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Functional Tests Scale Fidelity ...
Страница 536: ...536 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Software Software Updates ...
Страница 541: ...Index Index 541 Y yellow LED 48 51 yig filter see A12 YTF YTF see A12 YTF ...