Keysight NFA Series Noise Figure Analyzers Service Guide
455
Assembly Replacement Procedures
Input Connector Assembly
Input Connector Assembly
Removal
1.
Remove the instrument outer case. Refer to the
removal procedure.
2.
Remove the Front Frame Assembly. Refer to the
removal procedure.
3.
. Disconnect the semi-rigid cable W10 from the Input
Connector Assembly
(2)
.
Figure 15-50
Input Connector Assembly Removal
4.
Remove the two screws
(1)
that attach the Input Connector Assembly
(2)
to the chassis. The connector assembly can now be lifted from the
chassis.
5.
To remove the gusset
(5)
remove the three screws
(4)
, one on the side of
the chassis and two on the bottom.
Содержание N8973B
Страница 4: ...4 ...
Страница 44: ...44 Keysight NFA Series Noise Figure Analyzers Service Guide Overview How to Return Your Instrument for Service ...
Страница 132: ...132 Keysight NFA Series Noise Figure Analyzers Service Guide Instrument Messages Condition Messages ...
Страница 228: ...228 Keysight NFA Series Noise Figure Analyzers Service Guide RF Section Troubleshooting N8976B Troubleshooting ...
Страница 268: ...268 Keysight NFA Series Noise Figure Analyzers Service Guide Analog Digital IF Troubleshooting 25 MHz BW IF Section ...
Страница 322: ...322 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 332: ...332 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams NFA Computer Block Diagram ...
Страница 350: ...350 Keysight NFA Series Noise Figure Analyzers Service Guide Service and Diagnostics Menus Front Panel Key Test ...
Страница 470: ...470 Keysight NFA Series Noise Figure Analyzers Service Guide Assembly Replacement Procedures Front Frame Assembly ...