Keysight NFA Series Noise Figure Analyzers Service Guide
411
Assembly Replacement Procedures
RF Area (N8976B)
Low Band Switch
Removal
1.
. Remove the ribbon cable W18.
2.
Remove the semi-rigid cables W3, W8, and W9 using the 5/16 inch
wrench.
3.
Remove the two screws
(2)
using the T-10 driver. The low band switch
and bracket
(1)
can now be removed from the chassis.
Figure 15-17
Low Band Switch Cable Removal