130
Keysight NFA Series Noise Figure Analyzers Service Guide
Instrument Messages
Condition Messages
Condition Errors 801 to 899, Temperature
This series of errors corresponds to the bits in the
STATus:QUEStionable:TEMPerature
register. The second column in the
table shows the corresponding bit in that register.
An event with the error number shown in the table means the condition has
been detected. When the condition is cleared, an event with the error number
plus 1000 is generated. These error numbers can be viewed in the Show Errors
screen, along with the DETECTED and CLEARED indicators.
For example, error 801 indicates that the Ref Osc Oven Cold condition has
been detected; error 1801 indicates that condition has been cleared.
This register is summarized as bit 4 of the
STATus:QUEStionable
register, as
described in the section
“Condition Errors 601 to 699, Error Summaries” on
.
715
7
unused
717
8
Paused
status bit
only
719
9
Source Sweeping
status bit
only
The “Source Sweeping” bit is used to indicate
various conditions, depending on the Mode of
operation:
In the List Sequencer mode, it is used to indicate
that the sequencer is running
In the EMI Receiver mode, it is used to indicate
that scan results are available from the Frequency
Scan measurement
721
10
DC Coupled
W
723
11
unused
725
12
Source Waiting for
Trigger
status bit
only
727
13
unused
729
14
unused
Err#
Bit in
status
register
Message
Error or
Warning
More Information
Err#
Bit in
status
register
Message
Error or
Warning
More Information
801
0
Reference Oscillator Oven
Cold
W
(not currently in use)
Содержание N8973B
Страница 4: ...4 ...
Страница 44: ...44 Keysight NFA Series Noise Figure Analyzers Service Guide Overview How to Return Your Instrument for Service ...
Страница 132: ...132 Keysight NFA Series Noise Figure Analyzers Service Guide Instrument Messages Condition Messages ...
Страница 228: ...228 Keysight NFA Series Noise Figure Analyzers Service Guide RF Section Troubleshooting N8976B Troubleshooting ...
Страница 268: ...268 Keysight NFA Series Noise Figure Analyzers Service Guide Analog Digital IF Troubleshooting 25 MHz BW IF Section ...
Страница 322: ...322 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 332: ...332 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams NFA Computer Block Diagram ...
Страница 350: ...350 Keysight NFA Series Noise Figure Analyzers Service Guide Service and Diagnostics Menus Front Panel Key Test ...
Страница 470: ...470 Keysight NFA Series Noise Figure Analyzers Service Guide Assembly Replacement Procedures Front Frame Assembly ...