Appendix B
Verification and Calibration
96
Series N5700 User’s Guide
Verification
Verification procedures verify that the power supply is operating
normally and is within published specifications. There are two types
of verification tests:
Performance
These tests verify that the power supply meets all of the specifications listed
in Appendix A. They can also be used to verify that the power supply is
properly calibrated.
Calibration
These procedures calibrate the power supply.
If the power supply fails any of the tests or if abnormal test results
are obtained, try calibrating the unit. If calibration is unsuccessful,
return the unit to a Keysight Technologies repair facility (see
Appendix D).
Equipment Required
The equipment listed in the following table, or the equivalent to this
equipment, is required for the calibration and performance tests. A
test record sheet may be found at the back of this section.
Type
Specifications
Recommended Model
Digital Voltmeter
Resolution: 10 nV @ 1V;
Readout: 8 1/2 digits
Accuracy: 20 ppm
Keysight 3458A or equivalent
Current Monitor
15 A (0.1
Ω
) 0.04%, TC=5ppm/
°
C
100 A (0.01
Ω
) 0.04%, TC=5ppm/
°
C
300 A (0.001
Ω
) 0.04%, TC=5ppm/
°
C
Guildline 9230/15
Guildline 9230/100
Guildline 9230/300
Load Resistor
For 750 W models:
0.06
Ω
, 0.09
Ω
, 0.21
Ω
, 0.53
Ω
, 1.20
Ω
, 2.11
Ω
, 4.80
Ω
, 8.42
Ω
,
13.33
Ω
, 30.0
Ω
, 120
Ω
, 462
Ω
- all resistors 1kW minimum.
For 1500 W models:
0.03
Ω
, 0.04
Ω
, 0.10
Ω
, 0.26
Ω
, 0.60
Ω
, 1.05
Ω
, 2.40
Ω
, 4.21
Ω
,
6.67
Ω
, 15.0
Ω
, 60
Ω
, 231
Ω
- all resistors 2kW minimum.
Electronic Load
150 V, 100 A minimum (for Models N5741- N5750A)
150 V, 180 A minimum (for Models N5761- N5770A)
Keysight N3300A mainframe,
with 3 - N3305A modules
GPIB Controller
Full GPIB capabilities (for calibrating the unit over the GPIB) Keysight 82350B or
equivalent
Oscilloscope
Sensitivity: 1 mV
Bandwidth Limit: 20 MHz
Probe: 1:1 with RF tip
Keysight Infiniium or
equivalent
RMS Voltmeter
True RMS
Bandwidth: 20 MHz
Sensitivity: 100
µ
V
Rhode and Schwartz Model
URE3 or equivalent
Differential Amplifier Bandwidth: 20 MHz
LeCroy DA1855A, DA1850A,
or equivalent
Differential Probe
100:1/10:1 selectable
LeCroy DXC100A or eqivalent
Terminations
1 – 50
Ω
BNC termination
Variable-voltage
xfmr or AC source
Adjustable to highest rated input voltage range.
Power: 2000 VA