![Keysight N5247-60116 Скачать руководство пользователя страница 39](http://html.mh-extra.com/html/keysight/n5247-60116/n5247-60116_installation-note_1961527039.webp)
Installation Note N5247-90116 39
Description of the Upgrade
Installation Procedure for the Upgrade
Step 21. Remove Option 028 License
Procedure Requirements
— The analyzer must be powered up and operating to perform this procedure.
— The Network Analyzer program must
not
be running.
— A keyboard and mouse must be connected to the network analyzer.
Option 028 License Removal Procedure
1.
To start the Keysight License Manager, press
Start > Keysight License
Manager > Keysight License Manager
. A Keysight License Manager dialog
box will appear.
2.
Right click the on the desired option and click
Delete
.
3.
In the Select Desired Option list, click
028
.
4.
In the Keysight License Manager dialog box that appears, press or click
Yes
to confirm delete.
5.
A message displays stating that the option removal was successful.
Step 22. Enable Option 029
Procedure Requirements
— The analyzer must be powered up and operating to perform this procedure.
— The Network Analyzer program must
not
be running.
— Refer to the license message you received from Keysight: Verify that the
analyzer’s model and serial numbers match those on the license message
you received from Keysight.
— A keyboard must be connected to the network analyzer.
If Option 28 is not loaded on your PNA, proceed to
.
For this step, you will need a USB flash drive.
A single license file may contain more than one feature.