58 Installation Note N5247-90113
Installation Procedure for the Upgrade
-
Step 24. Reinstall the Front Panel Assembly
Refer to
for this step of the procedure. New parts are listed in
1.
Reconnect the ribbon cable (item
) to the A1 front panel interface board.
2.
Slide the front panel over the front-panel connectors.
3.
With a T-10 TORX driver, reinstall the 12 screws (item
) in the sides of
the frame.
— Be sure to install the two new screws (0515-1946) in the front panel, next to
test ports 3 and 4. Torque these screws to 9 in-lbs.
Figure 31
Front Panel Assembly Re-installation
Before installing the front panel assembly onto the analyzer, lift and
support the front of the analyzer chassis.