Keysight N5224B/25B Service Guide 3-9
Tests and Adjustments
About System Verification and Performance Tests
Certificate of Calibration
Keysight Technologies will issue a certificate of calibration upon successful
completion of system verification or completion of the performance tests. The
certificate of calibration will apply to the “system” (analyzer, calibration kit, test
cables, and any necessary adapters) if the system verification procedure is
used to confirm the system specifications. If the performance tests are used to
confirm instrument specifications, the certificate of calibration will apply to the
PNA as an independent instrument. The equipment and measurement
standards used for the tests must be certified and must be traceable to
recognized standards.
If you have a measurement application that does not use all of the
measurement capabilities of the analyzer, you may ask your local Keysight
Technologies service office to verify only a subset of the specifications.
However, this “limited calibration” creates the possibility of making
inaccurate measurements if you then use the analyzer in an application
requiring additional capabilities.
Содержание N5224B
Страница 3: ...3...
Страница 4: ...4...
Страница 12: ...12 Keysight N5224B 25B Service Guide Contents...
Страница 84: ...3 52 Keysight N5224B 25B Service Guide Tests and Adjustments Adjustments 3...
Страница 145: ...Keysight N5224B 25B Service Guide 4 61 Troubleshooting This page intentionally left blank...
Страница 212: ...5 50 Keysight N5224B 25B Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 17 4 Port Receiver Group...
Страница 213: ...Keysight N5224B 25B Service Guide 5 51 Theory of Operation Receiver Group Operation Figure 5 18 2 Port Receiver Group...
Страница 220: ...5 58 Keysight N5224B 25B Service Guide Theory of Operation Power Supply Group Operation 5...
Страница 225: ...Keysight N5224B 25B Service Guide 6 5 Replaceable Parts Rebuilt Exchange Assemblies Figure 6 1 Module Exchange Procedure...
Страница 490: ...6 270 Keysight N5224B 25B Service Guide Replaceable Parts Replaceable Parts Listings 6...
Страница 608: ...8 102 Keysight N5224B 25B Service Guide General Purpose Maintenance Procedures Correction Constants 8...
Страница 612: ...Appendix 4 Service Guide EEPROM Address Assignments and Location N5224 5A B PNA and N5244 5A B PNA X Instruments...
Страница 613: ...Installation Note Xxxxx xxxxx 3...