Installation Note N5222-90109
7
Step 1. Verify the Option, Serial, and New Model Numbers
If you haven’t done so already, follow the instructions in
“License Key Redemption” on page 4
must use the online Keysight HostID utility to get a new Host ID.
Refer to the license message you received from Keysight:
• Verify that the analyzer’s option number (626), serial number, and upgraded model number (N5222A) match
those on the license message you received from Keysight.
• Note the 12-digit license key for later use.
If the model number, serial number, or option number do not match those on the license message you received
from Keysight, you will not be able to install the option. If this is the case, contact Keysight for assistance.
“Getting Assistance from Keysight” on page 3
Step 2. Enter the New Model Number
1. Click
System > Service > Option Enable
.
2. Click
Repair.
3. Click the
Check if model shown is incorrect
check box.
4. Click
Begin Repair
.
5. In the drop down list that appears, select your new, upgraded PNA model number–N5222A.
6. Using the keyboard, enter the 12-digit license key in the box provided. The license key is printed on the
license message you received from Keysight. Enter this key
exactly
as it is printed on the message.
7. Click
Change Model
.
8. Click
Yes
in answer to the displayed question in the
Restart Analyzer?
box.
9. When the installation is complete, click
Exit
.
NOTE
The PNA contains adjustment data for the original frequency range only. Performance above
this range is not calibrated and may cause errors to be displayed. This will be corrected when
the source and receiver adjustments are performed.
NOTE
Previously installed software options require new license keys. Contact Keysight to receive
these keys.