Keysight N4372E Lightwave Component Analyzer User’s Guide
123
Maintenance
7
3 Take a new, dry soft-tissue and remove the alcohol, dissolved sediment
and dust, by rubbing gently over the surface using a small circular
movement.
4 Blow away any remaining lint with compressed air.
An Alternative Procedure
A better, more gentle, but more expensive cleaning procedure is to use an
ultrasonic bath with isopropyl alcohol.
1 Hold the tip of the connector in the bath for at least three minutes.
2 Take a new, dry soft-tissue and remove the alcohol, dissolved sediment
and dust, by rubbing gently over the surface using a small circular
movement.
3 Blow away any remaining lint with compressed air.
How to clean connector interfaces
Preferred Procedure
Use the following procedure on most occasions.
1 Clean the interface by pushing and pulling a new, dry pipe-cleaner into
the opening. Rotate the pipe-cleaner slowly as you do this.
2 Blow away any remaining lint with compressed air.
Procedure for Stubborn Dirt
Use this procedure particularly when there is greasy dirt on the interface:
1 Moisten a new pipe-cleaner with isopropyl alcohol.
2 Clean the interface by pushing and pulling the pipe-cleaner into the
opening. Rotate the pipe-cleaner slowly as you do this.
3 Moisten a new cotton-swab with isopropyl alcohol.
4 Using a new, dry pipe-cleaner, and a new, dry cotton-swab remove the
alcohol, any dissolved sediment and dust.
5 Blow away any remaining lint with compressed air.
CAUTION
Be careful when using pipe-cleaners, as the core and the bristles of the
pipe-cleaner are hard and can damage the interface.
Содержание N4372E
Страница 1: ...User s Guide Keysight N4372E Lightwave Component Analyzer ...
Страница 8: ......
Страница 22: ......
Страница 30: ...30 Keysight N4372E Lightwave Component Analyzer User s Guide 2 Setting Up the Instrument ...
Страница 92: ......
Страница 130: ......
Страница 136: ......
Страница 137: ...Keysight N4372E Lightwave Component Analyzer User s Guide 10 Definition of Terms Definition of Terms 138 ...