Keysight N2820/1A High-Sensitivity Current Probes
User’s Guide
7 Performance Verification
Average Gain of Zoomed-In Channel / 97
Average Gain of Zoomed-Out Channel / 101
Performance Test Record / 105
This chapter describes how to verify the performance of the N2820A and N2821A
probes. The performance verification requires an N2825A probe head and consists
of verifying the probe’s DC gain.
NOTE
The Average Gain verified in this chapter is listed as Gain in
NOTE
The performance measured in this procedure is for the probe by itself. Keysight high
performance real-time scopes apply probe correction that will further enhance the
performance of the probes.
NOTE
Keysight recommends a test interval of one year or 2000 hours of operation.
CAUTION
Electrostatic discharge (ESD) can quickly and imperceptibly damage or destroy high
performance probes, resulting in costly repairs. Always wear a wrist strap when handling
probe components and insure that cables are discharged before being connected.
Содержание N2820A
Страница 1: ...Keysight N2820 1A High Sensitivity Current Probes User s Guide...
Страница 30: ...30 N2820 1A User s Guide 2 Making Measurements Quick Start Figure 10 R Sense Entry Box...
Страница 74: ...74 N2820 1A User s Guide 4 Probe Calibration To Calibrate the Probe on InfiniiVision Oscilloscopes...
Страница 84: ...84 N2820 1A User s Guide 5 Characteristics and Specifications Dimensions...
Страница 94: ...94 N2820 1A User s Guide 6 Performance Data Plots Input Load Model...
Страница 106: ...106 N2820 1A User s Guide 7 Performance Verification Performance Test Record...
Страница 109: ...N2820 1A User s Guide 109...