66
N2820/1A User’s Guide
4
Probe Calibration
To Calibrate the Probe on Infiniium Oscilloscopes
To Calibrate the Probe on Infiniium Oscilloscopes
During this calibration, you may be prompted to make one or more
adjustments. These adjustments, located on the underside of the
probe amplifier, are shown in
Table 8
Required Calibration Test Equipment
Procedure
1
Turn the oscilloscope on.
2
Solder the sensor leads onto the N2825A user-defined head as
explained in
3
Attach the N2825A user-defined head to the probe.
Test Equipment
Description
N2825A User-Defined
Probe Head
Provided with probe.
Pomona 3788 BNC
Female to Minigrabber
Test Clips Breakout
http://www.pomonaelectronics.com/
for more information and how to order
this product.
Adapters
SMA, BNC, BNC Tee, BNC cables and cable clips as needed.
Screwdriver
For making adjustments during the procedure. This tool is
provided with the probe.
NOTE
Redirecting the Probe Comp Signal
On 9000-series and S-series Infiniium oscilloscopes, you can redirect
the Probe Comp signal to the Aux Output BNC connector on the rear
panel of the oscilloscope. This connector can then be connected to the
Pomona 3788 using a BNC cable. Click
Utilities
>
Calibration
and select
Probe Comp
in the
Aux Output
drop-down list.
Содержание N2820A
Страница 1: ...Keysight N2820 1A High Sensitivity Current Probes User s Guide...
Страница 30: ...30 N2820 1A User s Guide 2 Making Measurements Quick Start Figure 10 R Sense Entry Box...
Страница 74: ...74 N2820 1A User s Guide 4 Probe Calibration To Calibrate the Probe on InfiniiVision Oscilloscopes...
Страница 84: ...84 N2820 1A User s Guide 5 Characteristics and Specifications Dimensions...
Страница 94: ...94 N2820 1A User s Guide 6 Performance Data Plots Input Load Model...
Страница 106: ...106 N2820 1A User s Guide 7 Performance Verification Performance Test Record...
Страница 109: ...N2820 1A User s Guide 109...
Страница 110: ...This information is subject to change without notice Keysight Technologies 2018 March 2018 N2820 97006 www keysight com...