20
N2795/6A User’s Guide
1
Using the Probes
Functional Test
Use the following generic measurement procedure to ensure that
your probe is functioning properly.
1
Connect the probe to an oscilloscope channel input and ensure
the input impedance of the oscilloscope matches the output
impedance of the probe (50
W
).
2
Connect a grounding accessory to the probe and connect the
ground to the ground terminal on the oscilloscope.
3
Connect the probe tip to the oscilloscope’s probe compensation
output.
4
Enable autoscale on the oscilloscope.
5
A square wave should now be displayed on the oscilloscope, if
the probe is functioning properly.
Содержание N2795/6A
Страница 1: ...Keysight N2795 6A Single Ended Probes User s Guide...
Страница 4: ...4 N2795 6A User s Guide Contents...
Страница 29: ...Characteristics and Specifications 2 N2795 6A User s Guide 29 Dimensions...
Страница 30: ...30 N2795 6A User s Guide 2 Characteristics and Specifications...
Страница 34: ...34 N2795 6A User s Guide 3 Performance Data Plots...
Страница 48: ...48 N2795 6A User s Guide 4 Performance Verification...
Страница 50: ...Index 50 N2795 6A User s Guide...