Performance Verification
N2780/1/2/3B User’s Guide
45
4
Clamp the probe around the coil and ensure that the probe is as perpendicular
as possible to the coils of wire.
5
Set up the signal generator to the following settings:
Wave shape: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sine
Frequency: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 Hz
Amplitude:
N2780B: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6.5A
rms
N2781B: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3A
rms
N2782B: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6A
rms
N2783B: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6A
rms
Offset: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0.0 Vdc
6
Using the DMM, measure the voltage output of the probe and record this value
as the
DMM Voltage
7
Calculate the current by multiplying the
DMM Voltage
by the value in the
Multiplier
column. Record the result in the
Calculated Amperes
column.
8
Compare the
Calculated Amperes
to the
Passing Range
values to determine if the
probe passes the test.
Table 8
Test Results
Model Number
DMM Voltage
Multiplier
Calculated Amperes
Passing Range
N2780B
100
505.5A to 494.5A
N2781B
100
151.6A to 148.4A
N2782B
10
30.31A to 29.69A
N2783B
10
30.31A to 29.69A