 Frequency Compensation Adjustment

9

Frequency Compensation Adjustment

The N2771B High Voltage Probe has a Frequency Compensation 

adjustment that compensates for the input capacitance of the 

oscilloscope being used with the probe. Use a square-wave 

generator set for approximately 100 Hz output. Proceed with the 

following steps for frequency compensation.

1

Connect the probe to the oscilloscope.

2

Connect probe tip to square-wave generator.

3

Adjust the square wave generator for approximately 10 volts 

amplitude.

4

Adjust the oscilloscope time base for 20 

μ

s/div.

5

Use a trimmer tool to adjust the compensation capacitor for the 

flattest pulse top. See the figure below for the location of the 

adjustment.

Figure 1

Location of Frequency Compensation Adjustment

Cleaning

Clean only the exterior probe body and cables. Use a soft cotton 

cloth lightly moistened with a mild solution of detergent and water. 

Do not allow any portion of the probe to be submerged at any time.

WARNING

Dry the probe thoroughly before attempting to make voltage 

measurements.

CAUTION

Do not subject the probe to solvents or solvent fumes as these can cause 

deterioration of the probe body and cables.

Содержание N2771B

Страница 1: ...Keysight N2771B 15 kV High Voltage Probe User s Guide...

Страница 2: ...odify distribute or disclose the Software The EULA and the license set forth therein does not require or permit among other things that Keysight 1 Furnish technical information related to commercial c...

Страница 3: ...and nominal input capacitance between 6 and 20 pF The N2771B is a 1000 1 divider which extends the voltage measurement capability to maximum 15 kV DC AC peak 10 kVrms AC or 30 kVpeak 20 ns pulse widt...

Страница 4: ...ory I CAT I only Do not use the probe for measurements performed on circuits defined by any other measurement category or transient overvoltages of more than 1500V Refer to IEC Measurement Category De...

Страница 5: ...nnecting the probe WARNING The probe body should be kept clean and free of any conductive contamination Refer to the section on cleaning WARNING The ground lead is critical to the safe operation of th...

Страница 6: ...II is for measurements performed on circuits directly connected to the low voltage installation Example Household appliances portable tools and similar equipment Measurement Category III CAT III Measu...

Страница 7: ...r than the oscilloscope waveform if the probe attenuation on the oscilloscope has been set to 1 1 8 Turn off the high voltage source 9 Disconnect the Keysight N2771B High Voltage Probe from the high v...

Страница 8: ...8 kV DC or 5 kV AC rms Keysight recommends that an interval of 5 minutes be taken between measurements WARNING This high voltage probe is designed for use within Measurement Category I CAT I only Do...

Страница 9: ...imately 10 volts amplitude 4 Adjust the oscilloscope time base for 20 s div 5 Use a trimmer tool to adjust the compensation capacitor for the flattest pulse top See the figure below for the location o...

Страница 10: ...tio For input capacitance of 6 pF to 20 pF 1000 1 Bandwidth Designed For Use In 50 MHz 3 dB POLLUTION DEGREE 2 Input Resistance Operating Temperature 100 M when terminated by 1 M 0 C to 50 C Cable Len...

Страница 11: ...Voltage Derating 11 Voltage Derating The following graph shows the voltage versus frequency derating curve for the Keysight N2771B High Voltage Probe Figure 2 Voltage Versus Frequency Derating Curve...

Страница 12: ...12 Contacting Keysight Contacting Keysight Information on contacting Keysight Technologies can be found at www keysight com find contactus...

Отзывы: