Two-Oscilloscope System

2

Keysight UXR-Series MultiScope Hardware Configuration Guide

13

 

Connecting the Equipment

Parts in the N2164A (Two in Rack) Expansion Kit

In the MBASE expansion kit, these parts are used to connect the oscilloscopes in 

the MultiScope system:

These parts are used to create a long cable for performing channel deskew:

Part number

Qty

Description

54964-61632

5

1 m phase-stable cable

54964-60401

1

6-way splitter assembly with rack mount shelf

1250-3745

4

50

Ω

 SMA terminator

Part number

Qty

Description

54964-61632

2

1 m phase-stable cable

1250-2818

2

Right-angle male to female SMA adapter

1250-3758

1

Female/female SMA adapter (connector saver) 3.5 mm (f)

1250-3347

1

Push-on male SMA to female SMA adapter

Содержание MultiScope UXR Series

Страница 1: ...Keysight UXR Series MultiScope Hardware Configuration Guide ...

Страница 2: ...S government customers under its standard commercial license which is embodied in its End User License Agreement EULA a copy of which can be found at www keysight com find sweula The license set forth in the EULA represents the exclusive authority by which the U S government may use modify distribute or disclose the Software The EULA and the license set forth therein does not require or permit amo...

Страница 3: ...ning the necessary cables adapters and power dividers are required N2164A UXR Series Multi Frame Base Kit First Two Frames N2165A UXR Series Multi Frame Expansion Kit Add One Frame The N2164A and N2165A rack mount expansion accessories are set up so that you can expand your system one oscilloscope at a time NOTE The Infiniium UXR Series oscilloscope software supports up to 40 channels in a MultiSc...

Страница 4: ... an external host computer MultiScope software is included as a standard feature This software allows time synchronized acquisitions for systems of up to ten oscilloscopes with very low oscilloscope to oscilloscope jitter and drift This software can be used to configure capture and store real time voltage signals Not Included MultiScope systems also require Equipment rack Rack mount kits N2156A fo...

Страница 5: ...n in the following table After you have set up and connected the MultiScope system see Chapter 4 Next Steps starting on page 23 See Also The Infiniium UXR Series oscilloscope software s online help for information on using the MultiScope Hosted Instruments software of channels required 4 ch oscilloscopes of channels required 2 ch oscilloscopes Configuration required See 5 8 3 4 2 oscilloscope conf...

Страница 6: ...6 Keysight UXR Series MultiScope Hardware Configuration Guide ...

Страница 7: ...n Kit 13 Connecting the Oscilloscopes 14 3 Three to Five Oscilloscope Systems Expansion Equipment Required 18 Mounting Oscilloscopes in Rack 19 Connecting the Equipment 20 Parts in the N2165A Add One Expansion Kit 20 Connecting the Follower 2 Through Follower 4 Oscilloscopes 21 4 Next Steps Calibrate the Oscilloscopes in the MultiScope System 23 Use MultiScope Control Software 24 Infiniium Oscillo...

Страница 8: ...8 Keysight UXR Series MultiScope Hardware Configuration Guide ...

Страница 9: ...illoscopes environmental and rack Oscilloscopes are sensitive measurement instruments and some environmental conditions must be present to ensure that the oscilloscope delivers the best possible measurement accuracy Refer to the Infiniium UXR Series Real Time Oscilloscopes User s Guide for information on power requirements airflow cooling requirements and rack cabinet requirements ...

Страница 10: ...10 Keysight UXR Series MultiScope Hardware Configuration Guide 1 UXR Series MultiScope System Requirements ...

Страница 11: ... Equipment Required Table 1 Equipment Required for Two Oscilloscope Rack Mount System Model Description Quantity N2164A UXR Series Multi Frame Base Kit First Two Frames 1 N2156A Rack mount kit 2 Figure 2 Expansion Accessories Required for Two Oscilloscope Rack Mount System UXR Series oscilloscope UXR Series oscilloscope N2164A 2 oscilloscope kit for rack Rack 2 oscilloscope rack mount system ...

Страница 12: ... Rack Mount Kit User s Guide mount oscilloscopes in their designated rack locations NOTE Locate oscilloscopes in the rack so that the overall rack center of gravity is about half way up the rack Because of cable lengths use the recommended power divider location relative to the Leader and Follower oscilloscope locations Figure 3 Oscilloscope Locations in Rack Leader Oscilloscope 7U Top of Rack Fol...

Страница 13: ...m These parts are used to create a long cable for performing channel deskew Part number Qty Description 54964 61632 5 1 m phase stable cable 54964 60401 1 6 way splitter assembly with rack mount shelf 1250 3745 4 50 Ω SMA terminator Part number Qty Description 54964 61632 2 1 m phase stable cable 1250 2818 2 Right angle male to female SMA adapter 1250 3758 1 Female female SMA adapter connector sav...

Страница 14: ...puts 3 Connect one 54964 61632 cable from the Leader Sample_CLK_Out output to the power divider input 4 Connect two 54964 61632 cables between the power divider assembly outputs and the oscilloscopes Sample_CLK_In inputs 5 Connect one 54964 61632 cable between the Leader Trig Out and Follower 1 Aux Trig In connectors 6 Connect one 54964 61632 cable between the Leader Reference CLK Out and Follower...

Страница 15: ...re Configuration Guide 15 7 Connect the oscilloscopes to the Leader oscilloscope or external host computer through the desired remote interface connection LAN USB etc 8 If these are the last oscilloscopes to be connected see Chapter 4 Next Steps starting on page 23 ...

Страница 16: ...16 Keysight UXR Series MultiScope Hardware Configuration Guide 2 Two Oscilloscope System ...

Страница 17: ...17 Keysight UXR Series MultiScope Hardware Configuration Guide 3 Three to Five Oscilloscope Systems Expansion Equipment Required 18 Mounting Oscilloscopes in Rack 19 Connecting the Equipment 20 ...

Страница 18: ... Frame of scopes 2 N2156A Rack mount kit of scopes Figure 4 Expansion Accessories Required for Three to Five Oscilloscope Rack Mount System UXR Series oscilloscope UXR Series oscilloscope N2164A 2 oscilloscope kit for rack UXR Series oscilloscope N2165A 1 oscilloscope expansion kit UXR Series oscilloscope N2165A 1 oscilloscope expansion kit UXR Series oscilloscope N2165A 1 oscilloscope expansion k...

Страница 19: ...oscilloscopes in their designated rack locations NOTE Locate oscilloscopes in the rack so that the overall rack center of gravity is about half way up the rack Because of cable lengths use the recommended power divider location relative to the Leader and Follower oscilloscope locations Figure 5 Oscilloscope Locations in Rack Top of Rack Leader Oscilloscope 7U Follower 1 Oscilloscope 7U Follower 2 ...

Страница 20: ... Series MultiScope Hardware Configuration Guide 3 Three to Five Oscilloscope Systems Connecting the Equipment Parts in the N2165A Add One Expansion Kit Part number Qty Description 54964 61632 3 1 m phase stable cable ...

Страница 21: ... Follower 4 Power divider Place 1250 3745 terminators on all unused divider outputs Connections Back Reference CLK In Out Trig Out Aux Trig In Sample CLK In Out Reference CLK In Out Trig Out Aux Trig In Sample CLK In Out Reference CLK In Out Trig Out Aux Trig In Sample CLK In Out Reference CLK In Out Trig Out Aux Trig In Sample CLK In Out Reference CLK In Out Trig Out Aux Trig In Sample CLK In Out...

Страница 22: ...e power divider output that will be used for the oscilloscope 2 Connect one 54964 61632 cable between the power divider output and the oscilloscope s Sample_CLK_In input 3 Connect one 54964 61632 cable between the previous oscilloscope s Trig Out connector to this oscilloscope s Aux Trig In connector 4 Connect one 54964 61632 cable between the previous oscilloscope s Reference CLK Out connector to...

Страница 23: ...perating temperature of the oscilloscopes in the rack s or stackmount configuration once the entire system has reached a stable thermal condition The temperature deviation from the previous calibration is displayed in the Calibration dialog box of each oscilloscope If the inlet air temperature is not within the calibration temperature range the oscilloscope will provide a visual warning on the scr...

Страница 24: ...om the Infiniium Offline software you have only edge and glitch trigger modes No additional computer is required to run the Infiniium Offline software No space for an additional computer is necessary MultiScope software is included as a standard feature Refer to the Infiniium oscilloscope software s Online Help for instructions on setting up and controlling the MultiScope system In particular note...

Страница 25: ...nfiguration Guide 25 Host Computer Requirements For the D9010BSEO Infiniium Offline oscilloscope analysis software Keysight recommends a host computer with A Windows 10 64 bit operating system with at least 8 GBytes of RAM and 25 GBytes of hard drive space ...

Страница 26: ...lti Frame Base Kit The parts used to perform channel deskew are These parts are used to create a long cable that can be connected from Cal Out to each input channel in the MultiScope system Part number Qty Description 54964 61632 2 1 m phase stable cable 1250 2818 2 Right angle male to female SMA adapter 1250 3758 1 Female female SMA adapter 3 5 mm f 1250 3347 1 Push on male SMA to female SMA adap...

Страница 27: ...50 2818 right angle SMA adapter are included with the UXR Series oscilloscope UXR Series oscilloscope input connector type Adapter needed for channel deskew cable connection 3 5 mm 5061 5311 adapter 3 5 mm female to female 1 85 mm 5061 5311 adapter 3 5 mm female to female 1 0 mm 5067 1393 adapter 1 mm ruggedized female to 2 92 mm female ...

Страница 28: ...28 Keysight UXR Series MultiScope Hardware Configuration Guide 4 Next Steps ...

Страница 29: ...es 3 F final steps 23 follower oscilloscope 3 H host computer 4 host computer requirements 25 I in the box MultiScope control software 24 IO Libraries Suite 24 M MultiScope hardware configuration 3 MultiScope product overview 3 N N2156A rack mount kit 11 18 N2164A UXR Series Multi Frame Base Kit First Two Frames 3 11 13 18 N2165A UXR Series Multi Frame Expansion Kit Add One Frame 3 18 20 P parts i...

Страница 30: ...30 Keysight UXR Series MultiScope Hardware Configuration Guide Index ...

Отзывы: