89600 VSA Software Configuration

M9703B Hardware Extension of 89600 VSA Software

7

89600 VSA Software Configuration

Configuring the 89600 VSA software application to use the
digitizer hardware

Create a configuration for Keysight M9703B digitizer

1. Start the 89600 VSA software application.

2. Configure the digitizer as the VSA measurement input by creating an

'Analyzer

Configuration'

that uses your digitizer as the

Logical Instrument

(ADC) by per-

forming the following steps:

a. From the menu, go to Utilities > Hardware > Configurations, and click on

Add New Configuration button (

). The 'New Hardware Configuration'

dialog opens.

b. De-select 'Simulate Hardware' if it is selected (1).

c. Select Keysight MD2 digitizers(2) from the list of

'Possible Logical Instru-

ment'

and drag it or click on the

button (3) to add it to the

'Con-

figuration'

box .

Содержание M9703B

Страница 1: ...Configuration and Measurement Instructions Keysight M9703B High Speed Digitizers Hardware Extension of 89600 VSA Software...

Страница 2: ...and 27 405 3 and Department of Defense FAR Supplement DFARS 227 7202 the U S government acquires commercial computer software under the same terms by which the software is customarily provided to the...

Страница 3: ...Hz in interleaved mode SR1 F10 390 MHz 781 MHz in interleaved mode SR2 F05 625 MHz 650 MHz in interleaved mode SR2 F10 625 MHz 1 25 GHz in interleaved mode The present document aims to describe how to...

Страница 4: ...9703B product page www keysight com find M9703B Technical Support Drivers Firmware and Software Once downloaded execute MD2DigitizersHwExt_Setup versions exe and follow the installation process VSA 89...

Страница 5: ...n the MD2 Digitizer Hardware Extension has to be re installed There are two solutions Option 1 From Windows Control Panel Programs Programs and Features select and right click on Keysight MD2 Digitize...

Страница 6: ...the system In order to proceed execute C Program Files Keysight MD2 89600 VSA Connectivity MD2Digitizers 89600VSAIntegration exe or using your installation path Follow the installation process Click...

Страница 7: ...tizer as the VSA measurement input by creating an Analyzer Configuration that uses your digitizer as the Logical Instrument ADC by per forming the following steps a From the menu go to Utilities Hardw...

Страница 8: ...n list 4 e You may either use the default name or specify another name for this ana lyzer configuration 5 f Click OK to close the New Hardware Configuration dialog 6 g Before closing the Hardware Conf...

Страница 9: ...e sig nals please refer to the User manual of your digitizer section Multi Purpose Inputs and Outputs Enable external clock See External Clock page 12 Enable interleaving See Interleaved Mode page 11...

Страница 10: ...the option MD2 Auto Calibration from Input Extensions window then to manually start the calibration once The following calibrations will be automatic However the MD2 Auto Calibration mode is not reco...

Страница 11: ...Interleaving allows to double the sampling rate with the trade off of halving the available center frequency range as shown in the table below Model Option Mode Center Frequency Range ADC Sampling Rat...

Страница 12: ...own in the table below Models Option Frequency Range M9703B SR1 1 8 to 2 GHz SR2 1 8 to 3 2 GHz To use this option first connect an external clock signal in the accepted range to the CLK IN connector...

Страница 13: ...89600 VSA Software Configuration M9703B Hardware Extension of 89600 VSA Software 13...

Страница 14: ...is changed an internal calibration is required if this new configuration has not yet been calibrated in the current session A manual calibration can be performed at any time from the menu Utilities C...

Страница 15: ...izer inputs is Logical Channel Digitizer Input 1 IN1 2 IN3 3 IN5 4 IN7 5 IN2 6 IN4 7 IN6 8 IN8 From the menu Input Analog the following parameters can be set Range Thedigitizer offers several full sca...

Страница 16: ...s from the instantaneous analysis bandwidth IBW supported by the hardware digital down converter the down conversion is done by the hardware at the closest highest IBW and remaining processing is mana...

Страница 17: ...External Magnitude and Channel Delay available for External Magnitude and Channel Trigger Holdoff not supported by the digitizer Soft Front Panel Control The Disconnect feature can be used to pause t...

Страница 18: ...data recording from the digitizer to the PC s disk drive The length of waveform recording time is limited by the memory available on the digitizer Setup Save Recall The digitizer state is not saved w...

Страница 19: ...click on the button to add it to the Configuration box b Select Keysight MD2 digitizers from the list of Possible Logical Instrument and drag it or click on the button to add it to the Configuration...

Страница 20: ...select Input Channel Custom Use the Channel Configuration GUI to reconfigure the channel mapping from the physical hardware input ports to the VSA measurement channel inputs See picture below By defa...

Страница 21: ...s different with a single digitizer and in multi module mode The frequency reference can be accessed from Utilities Frequency reference Using a single digitizer the frequency reference is the internal...

Отзывы: