Keysight M9502A/M9505A AXIe Chassis User Guide
83
Chassis Synchronization and Triggering
Features and Functions
In the above example, TRIG 0 is enabled as an output, and will trigger when
slot 1 sends a STRIG signal. TRIG 2 is also enabled, and will trigger when an
external SMA trigger is applied.
STRIG
The STRIG signal is a Bidirectional Low Voltage Differential Signaling (BLVDS)
star trigger with the following characteristics:
–
Point-to-point and bidirectional between the ESM (logical slot 1) and
instrument slots (physical slots 1 through
n
)
–
Slot terminated with 100
W
differential and 50
W
series (optional)
–
All STRIG lines are trace length matched to 0.025")
Use When:
The STRIG bus’s trace matched lines ensure the lowest skew when
synchronously triggering multiple modules.
STRIG Bus
Slot 1
Slot n
Slot 2
BLVDS Buffer
BLVDS Buffer
BLVDS Buffer
Chassis Backplane
100W
ESM
100W
100W
50
50
50
50
50
50
BLVDS
Buffers
50
50
50
50
50
50
100W
100W
100W
Содержание M9502A
Страница 1: ...Keysight M9502A 2 Slot and M9505A 5 Slot AXIe Chassis User Guide...
Страница 2: ......
Страница 30: ...20 Keysight M9502A M9505A AXIe Chassis User Guide Introduction Product Warranty...
Страница 68: ...58 Keysight M9502A M9505A AXIe Chassis User Guide Navigating the Web Interface Chassis Health Page...
Страница 74: ...64 Keysight M9502A M9505A AXIe Chassis User Guide Using the Soft Front Panel SFP Screens...
Страница 82: ...72 Keysight M9502A M9505A AXIe Chassis User Guide Features and Functions Shelf Management Functions...
Страница 118: ...108 Keysight M9502A M9505A AXIe Chassis User Guide For Module Developers Chassis Backplane Connections...
Страница 138: ...128 Keysight M9502A M9505A AXIe Chassis User Guide Glossary...
Страница 141: ......