Step 4: Install the Modules

5. Follow the installer prompts. Choose a "Complete" installation to install all software and documentation, or

a "Custom" installation to select from a listing of modules and other features.

6. After installation is complete, power down the host PC, and then the chassis if you are using a remote con-

troller.

If you are using a remote controller, Shut Down the PC BEFORE you power down the chas-

sis. When you restore power, power up the chassis BEFORE you power up the PC.

Step 4: Install the Modules

PXI hardware does not support "hot-swap" (changing modules while power is applied to the chassis)

capabilities. Before installing or removing a module to/from the chassis, power off the chassis to pre-
vent damage to the module.

The Agilent M9302A PXI Local Oscillator, Agilent M9351A PXI Downconverter, Agilent M9360A
PXI Attenuator/Preselector, and modules are compatible with cPCI(J1), PXI-1, or PXIe hybrid slots.
The Agilent M9202A PXIe IF Digitizer module is compatible with a PXIe or PXIe hybrid slot.

1. Make sure that the line cord is plugged in to establish earth ground and that the chassis power switch is in

the Off (Standby) position.

2. If the chassis has multiple fan speed settings, ensure that the fans are set to automatic. Do not set the fan

speed to low or turn it off.

3. Position the chassis so that there is ample space between the chassis fan intake and exhaust vents. Block-

age by walls or obstructions affects the air flow needed for cooling. (Refer to the chassis documentation for
more information about cooling).

4. Before installing a module, be sure to remove the plastic thread protectors from the mounting screws.

Before inserting the module into the chassis, back the mounting screws out to assure that there is no inter-
ference between the screws and the mounting rails.

Agilent M9392A PXI Vector Signal Analyzer Startup Guide

13

Содержание M9392A

Страница 1: ...Startup Guide Keysight M9392A PXI Vector Signal Analyzer 50 MHz to 26 5 GHz...

Страница 2: ......

Страница 3: ...h the software is customarily provided to the public Accordingly Keysight provides the Software to U S government customers under its standard commercial license which is embodied in its End User Lice...

Страница 4: ...control and data I O connections are for connection to Category I sources unless otherwise marked or described in the user documentation Exercise extreme caution when a shock hazard is present Lethal...

Страница 5: ...for professional use and is for use in electromagnetic environments outside of the home A A This product complies with the WEEE Directive marketing requirement The affixed product label above indicat...

Страница 6: ...9351A 19 M9392A Block Diagram Reference Table for M9202A 19 Step 5 Verify Operation of the Agilent M9392A PXI Vector Signal Generator 20 Conduct M9392A Field Calibration 21 Required Hardware 21 Step 6...

Страница 7: ......

Страница 8: ...Documentation Map Documentation Map Agilent M9392A PXI Vector Signal Analyzer Startup Guide 7...

Страница 9: ...tup Guide and the documentation listed below are on the CD and at www agilent com find M9392A go to Document Library Manuals l M9392A Soft Front Panel help system l M9392A device driver API help syste...

Страница 10: ...unpacking a module inspect it for any shipping damage Report any damage to the shipping agent imme diately as such damage is not covered by the warranty see warranty information at beginning of this d...

Страница 11: ...up Guide in hard copy M9202A 1 9320 6741 China RoHS Addendum for Digitizer Products 1 9320 6698 China RoHS Addendum for Vector Signal Analyzer 1 5023 1450 Wrench socket extension 5 16 inch SMA 1 5002...

Страница 12: ...ay be required for operation than is required for installation The amount of space listed above is required for installation 4 Note NET Framework Runtime Components are installed by default with Windo...

Страница 13: ...This software must be installed first l If you wish to install the Agilent 89600 Vector Signal Analyzer Software do so AFTER installing the IOLS and BEFORE installing the drivers located on the Agilen...

Страница 14: ...PXI Attenuator Preselector and modules are compatible with cPCI J1 PXI 1 or PXIe hybrid slots The Agilent M9202A PXIe IF Digitizer module is compatible with a PXIe or PXIe hybrid slot 1 Make sure that...

Страница 15: ...front panel to the chassis using the two module front panel mounting screws Performance may suffer if the screws are not tightened properly 7 Verify that the PXI chassis fans are operable and free of...

Страница 16: ...emi rigid SMA SMA M9361A A3 M9361 20001 M9360A RF 2 OUT M9361A RF IN Cable semi rigid SMA SMA M9361A A4 M9361 20002 M9360A LO 2 OUT M9361A LO IN Cable semi rigid SMA SMA M9351A A5 M9351 20002 M9360A L...

Страница 17: ...B MMCX Cable Removal Tool to disconnect SMB and MMCX cables from the module front panel connectors To avoid damage to the cables or connectors pull the cable straight away from the connector Do not us...

Страница 18: ...ut level 0 dBm 3 dB 100 MHz REF 1 OUT Frequency 100 MHz M9351A Downconverter Output level 0 dBm 3 dB LO OUT Output frequency l Using M9351A Downconverter 3 0 GHz to 10 GHz 3 65 GHz higher than RF IN l...

Страница 19: ...o 160 dBm l Frequency 100 kHz to 26 5 GHz l RF IN to RF 1 OUT insertion loss 2 dB RF 2 OUT M9360A PXI Attenuator Preselector l RF 2 OUT 30 dBm to 160 dBm l Frequency o Bypassed 2 75 GHz to 26 5 GHz o...

Страница 20: ...MHz o IF OUT 500 MHz BW range 1 kHz to 40 MHz l Normal Downconverter path output is 2 dBm 500 mV p p o The RF IN to IF OUT gain is 38 dB l The BW is dependent on the RF IN Frequency o Lower side BW R...

Страница 21: ...duct a Self Test of the M9392A 1 Open the M9392A soft front panel SFP 2 Upon opening the SFP you are presented with the Connect to Instrument dialog Select all the modules that are components of the M...

Страница 22: ...Cable A Field Calibration should be performed upon initial setup and whenever one of more of the modules that compose the M9392A PXI Vector Signal Analyzer is removed reconfigured into different slot...

Страница 23: ...Program Files IVI Foundation IVI Drivers AgM9392 Example programs are provided to dem onstrate most driver functionality for example C Program Files IVI Foundation IVI Drivers AgM9392 Exa mples The re...

Страница 24: ...orqued to 8 Lb In 0 904 Nm l See Agilent M9392A Modular Instrument Connections on page 15 for instrument level cabling M9392A cable connections are also available from the M9392A SFP by selecting Conf...

Страница 25: ...supplies are functioning within their tolerances and all hardware is func tioning Indicates that all power supplies are functioning within their tolerances but there may be other hard ware failures l...

Страница 26: ...hout a 10 MHz Ref In LOCKED green with a 10 MHz Ref In l Ref LOCKED green l Temperature green l Voltage 12 5 3 3 12 green l Temperature green l Voltage 12 5 3 3 12 green l Ref Level OK green l Tempera...

Страница 27: ...good after reinstalling the other modules the module that was pulling the power supply down needs to be repaired iii If the voltage is bad after reinstalling the other modules troubleshoot the chassis...

Страница 28: ...lect Utility Reset 2 Select the Display tab a Set Display Frequency b Frequency Display Auto Scale unchecked c Scale 10 dB div d Ref Level 30 dBm 3 Under the Meas tab the Settings should be as follows...

Страница 29: ...ops approximately 10 dB each time the Input Power level is decreased ii The attenuator clicks as it switches The attenuator switches when stepping between the levels displayed in the table in Step 11...

Страница 30: ...age the downconverter Once the Input Level is reset and the OVERLOAD condition cleared Select Sweep Cont to restart the sweep 1 Select Utility Reset 2 Select the Display tab a Set Display Frequency b...

Страница 31: ...d c Scale 10 dB div d Ref Level 30 dBm 8 Under the Meas tab the Settings should be as follows a Use Preselector unchecked b Frequency 2 25 GHz c Input power set to 30 dBm d Bandwidth set to 250 MHz e...

Страница 32: ...level Measure the IF OUT levels from 2 25 GHz to 2 75 GHz 1 Source input frequency 2 25 GHz 2 Power level 10 dBm 3 M9392A SFP settings a Frequency 2 25 GHz b Bandwidth 250 MHz c Input power 10 dBm 4 A...

Страница 33: ...hin 3 dB c If power levels vary more than expected check that all cables are properly connected and torqued d Recheck areas of concern e If power levels still vary more than expected perform the M9361...

Страница 34: ...m 30 dBm 50 dBm 30 dBm 30 dBm 40 dBm 11 On the Meas Tab select Use Preselector The power level should be close to the same level when in the thru path Use Preselector unchecked and the preselector pat...

Страница 35: ...Preselector off 250 MHz 500 MHz when bandwidth is 50 MHz 468 MHz when bandwidth is 50 MHz 2 75 GHz to 26 5 GHz Preselector on 30 MHz 468 MHz 2 25 GHz to 2 75 GHz M9361A path 250 MHz 500 MHz Frequency...

Страница 36: ...respective driver help file from these program groups in the Win dows Start menu o IVI Driver Help Agilent IVI Drivers AgM9392 Documentation o LabVIEW Driver Help Agilent M9392 M9392 LabVIEW Help 3 U...

Страница 37: ...s installed in the chassis If you have a version older than 1 1 install the latest software from the supplied CD or from the Drivers Software tab on the Technical Support page at www agilent com find...

Страница 38: ...application again and load the license epr file sent to you by Agi lent Support see Step 2 above The license file update should only take several seconds 5 To verify the streaming upgrade is complete...

Страница 39: ...xternal RAID packages have been tested to guarantee sustained data rates with the M9392A These predefined mass storage packages ensure performance when used with the recommended controllers l Recommen...

Страница 40: ...Streaming Data Overview System diagram Agilent M9392A PXI Vector Signal Analyzer Startup Guide 39...

Страница 41: ...A firmware version 512 or greater If you need to upgrade your system install the latest software from the supplied CD or from the Drivers and Software tab on the Technical Support page at www agilent...

Страница 42: ...ndow in the Agilent 89600 VSA software by selecting Util ities Hardware Configurations from the menu bar When you hover over the last element in this sequence you will see a listing of the relevant co...

Страница 43: ...strument s to use for each logical instrument in the con figuration section c Name the new configuration in the 3 Name the configuration section The default name Analyzer1 may be changed Use a meaning...

Страница 44: ...and is highlighted Notice the red Hint in the fol lowing figure Since the VSA software has selected the modules in the lowest numbered left most slots it appears that the M9202A Adc is being used in b...

Страница 45: ...f the other dropdown choices select None c When you have configured this digitizer only logical instrument Name it appropriately 7 To create a two channel digitizer only configuration repeat Step 4 an...

Страница 46: ...procedure found in the M9392A SFP Help system Open the help system Pro grams Agilent M9392 M9392 Help and navigate to the Using Measurement Capabilities of the Agilent 89600 Series VSA Software topic...

Страница 47: ...ing the modules identified by their VISA address in the 2 Select the instrument s to use for each logical instrument in the con figuration section c Name the new configuration in the 3 Name the config...

Страница 48: ......