Coordinating Multiple PXIe Oscilloscope Modules
8
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
51
The Trigger Menu on the master oscilloscope module is used to define the trigger
condition. You can select from all the usual trigger types (Edge, Edge Then Edge,
Pulse Width, etc.). You are limited to master module input sources.
Step 4: (When Necessary) Connect Chassis Bus Segments
If the PXI oscilloscope modules are in different chassis trigger bus segments
(which are isolated by default), you must use the chassis Soft Front Panel (SFP)
application to connect the selected PXI_TRIG lines between trigger bus segments.
For example, if:
• The master module is in segment 1 and it is using PXI TRIG 0 as the trigger line.
• Two slave modules are in segment 3 and they will use PXI TRIG 1 and
PXI TRIG 2 as arm lines.
You would fllow these steps:
1
Click the Keysight IO Libraries Suite icon in the task bar and choose
Connection
Expert
from the popup menu.
2
In the Keysight Connection Expert window,
PXI/AXIe Chassis > Chassis Triggers
tab, hover over the lines in the bus segments you would like to reserve, and
then click
Click to reserve
.