Installation Note E8251-90396
21
cd "SPECIALS LATCH VALUE",13,+1.100000001
cd "SPECIALS LATCH VALUE",14,+4.300000001
cd "SPECIALS LATCH VALUE",15,+4.300000001
cd "SPECIALS LATCH VALUE",16,+2.700000001
cd "SPECIALS LATCH VALUE",17,+2.700000001
cc "SPECIALS LATCH VALUE"
ky 20622
Verify Signal Generator Calibration
1. Turn the signal generator on and allow it to warm up for 30 minutes.
2. Perform the following adjustment/performance tests to verify signal generator calibration.
Refer to the PSG Family Signal Generators Calibration Software:
NOTE
It is normal to see an “unlock” indicator prior to performing an adjustment.
Adjustment Tests
a. ADC Calibration
b. Frac-N Pretune Calibration
c. Offset Loop Pretune Calibration
d. Yig-Driver Pretune Calibration
e. KV vs. Frequency Calibration
f. FM Scale Offset Calibration
g. FM Path Offset Calibration
h. FM In-Band DCFM Calibration
i. FM Out-of-Band YO Calibration
Performance Tests
a. Self Tests
b. Harmonic Spurious
c. Sub-Harmonic Spurious
d. Non-Harmonic Spurious
e. Single-Sidedband Phase Noise