3 Input/Output Functions
Input/Output
Frequency Interpolation
This setting controls how the correction values per-bucket are calculated. We interpolate between
frequencies in either the logarithmic or linear scale.
This setting is handled and stored individually per correction set.
See
Key Path
Input/Output, Corrections, Properties
Remote Command
[:SENSe]:CORRection:CSET[1]|2|...|8:X:SPACing LINear | LOGarithmic
[:SENSe]:CORRection:CSET[1]|2|...|8:X:SPACing?
Example
CORR:CSET:X:SPAC LIN
Preset
Unaffected by a Preset. Set to Linear by Restore Input/Output Defaults.
State Saved
Saved in instrument state.
Initial S/W Revision
A.02.00
Modified at S/W Revision
A.14.00
Interpolation
For each bucket processed by the application, all of the correction factors at the frequency of interest
(center frequency of each bucket) are summed and added to the amplitude. All trace operations and post
processing treat this post-summation value as the true signal to use.
To effect this correction, the goal, for any particular start and stop frequency, is to build a correction trace,
whose number of points matches the current Sweep Points setting of the instrument, which will be used to
apply corrections on a bucket by bucket basis to the data traces.
For amplitudes that lie between two user specified frequency points, we interpolate to determine the
amplitude value. You may select either linear or logarithmic interpolation between the frequencies.
If we interpolate on a log scale, we assume that the line between the two points is a straight line on the log
scale. For example, let’s say the two points are (2,4) and (20,1). A straight line between them on a log scale
looks like:
104
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...