Confirm Required Equipment

Confirm Required Equipment

This section lists the minimum computer hardware, operating system, and software that are 

required to perform firmware upgrades on PSG/ESG signal generators.

Computer

IBM compatible PC with a 486 MHz processor or 

greater

RAM Required

64 MB or more

Hard Disk Space Required

20 MB or more

This could increase significantly if a large library of 

firmware upgrade files are expected to be maintained 

on the PC. A single firmware release consists of 

approximately 5 to 8 MB of data.

CD-ROM Drive or Modem

Use of a CD-ROM drive is only required when the 

computer being used does not have Internet access.

Use of a 56 kbps modem is only required for systems 

downloading firmware updated from the Internet.

RS-232, GPIB, or LAN Connection

— If you plan to perform a firmware upgrade using 

RS-232, refer to the “RS-232 Installation Kit (Only 

for use with E44xxB ESG Signal Generators)” on 

page 6.
While making the RS-232 connection, a 5-mm nut 

driver, part number 8710-1219, is used to remove 

the screws from the serial interface (f) to (f) adapter.

— If you plan to perform a firmware upgrade using 

GPIB, refer to “GPIB Interface Requirements” on 

page 6.

— If you plan to perform a firmware upgrade using 

LAN, a 10Base-T LAN interface is required. Contact 

your network administrator if problems are 

encountered.

Operating System

Windows XP, 2000, or Windows NT 4.0 (with Service 

Pack 5 or above)

Internet Browser

Use of an Internet Browser is only required for 

downloading firmware files from the Internet or for 

viewing firmware release information on your local PC 

or over the Internet.

Microsoft Internet Explorer 4.0 and above, or Netscape 

Navigator 4.0 and above

Either browser must be Java enabled with Java 1.02 or 

above.

Содержание E44 8C Series

Страница 1: ...Keysight ESG PSG Signal Generators E44xxB E44x8C E82x7C E82x7D Firmware Upgrade Guide ...

Страница 2: ...ied warranties of merchantability and fitness for a particular purpose Keysight Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing performance or use of this material Manual Part Number E4400 90549 Supersedes September 2013 Edition Edition 1 December 2015 Printed in USA Malaysia Published by Keysight Technologie...

Страница 3: ...ge 5 2 Confirm required equipment on page 6 3 Connect an RS 232 GPIB or LAN interface to a PC on page 8 4 Install PSG ESG Upgrade Assistant 5 12 and firmware files on page 11 5 Use the PSG ESG Upgrade Assistant 5 12 with firmware files on page 12 6 Check signal generator functionality on page 5 Product Affected All Serial Numbers All Options All except Option H98 To Be Performed By X Keysight Tech...

Страница 4: ...s new and previous versions of firmware installations to be performed using either RS 232 or GPIB on E44xxB ESG Signal Generators GPIB or LAN on E44x8C ESG Signal Generators GPIB or LAN on E824xA 5xA PSG Signal Generators GPIB or LAN on E82x7C D PSG Signal Generators it has an associated HTML page that describes the following information for each firmware upgrade version information about new feat...

Страница 5: ...e error queue appears in the display text area Refer to the signal generator error messages list on the documentation CD for information about each error message After resolving all problems causing errors press Clear Error Queue s If the ERR annunciator is off the signal generator functionality check has passed The signal generator must warm up for at least five minutes and all errors must be cle...

Страница 6: ...RS 232 GPIB or LAN Connection If you plan to perform a firmware upgrade using RS 232 refer to the RS 232 Installation Kit Only for use with E44xxB ESG Signal Generators on page 6 While making the RS 232 connection a 5 mm nut driver part number 8710 1219 is used to remove the screws from the serial interface f to f adapter If you plan to perform a firmware upgrade using GPIB refer to GPIB Interface...

Страница 7: ...e Upgrade CD ROM E4400 10501 1 RS 232 Cable Kit Serial RS 232 Cable 8120 6188 RS 232 Null Modem 5181 6639 Adapter f to f 1252 7825 E4400 60049 Interface Cards Manufacturer Driver Software Version Windows Version 82341C for ISA bus computers Keysight Technologies I O Libraries SICL VISA J 01 01 00 XP 2000 NT 4 0 82350A for PCI bus computers PCI GPIB National Instruments NI 488 2M 1 60 XP 2000 NT 4 ...

Страница 8: ...nnected to one of the COM ports on the PC using an RS 232 serial cable with a null modem connected between the COM port and the signal generator s RS 232 connector Refer to the RS 232 Installation Kit Only for use with E44xxB ESG Signal Generators on page 7 for more information 2 Set the RS 232 parameters as follows on the E44xxB ESG signal generators a Press the following keys Ensure that the sel...

Страница 9: ...o installing and configuring the GPIB interface card in your computer 2 Set GPIB parameters as follows on PSG ESG signal generators a Press the following keys for the model being set b Use the numeric keypad the arrow keys or rotate the front panel knob to set the desired address and press Enter The signal generator s GPIB address is set to 19 at the factory The acceptable range of addresses is 0 ...

Страница 10: ... generators a Press the following keys for the model being set b Use the numeric keypad to set the desired address and press Enter The signal generator LAN address is not set at the factory Contact your network administration to obtain an appropriate IP address c Set the Sockets SCPI parameters by pressing Utility GPIB RS 232 LAN LAN Services Setup Sockets SCPI On to use the E44x8C firmware instal...

Страница 11: ...rName Downloads The default installation folder can be changed using Save As from the Save drop down menu shown at the bottom of the browser 3 Download and install a version of PSG or ESG firmware files onto your computer The installation program places the selected firmware files in a default installation folder C Users userName Downloads The default installation folder can be changed using Save ...

Страница 12: ... values that are set on your signal generator For example the signal generator is usually set to GPIB Port 0 and Address 19 and the GPIB interface card in the computer is usually set to GPIB Port 0 and Address 21 These values are chosen by convention but any address in the range 0 30 inclusive may be used For further information refer to the documentation that came with your GPIB interface card If...

Страница 13: ...designated as New If an old version of firmware is to be installed select the Allow Old Revisions checkbox Note that installing old versions of firmware with new hardware may generate errors c Under Upgrade Selection verify that all firmware components are to be upgraded By default all upgrade components are selected If you want to perform a partial upgrade and do not already know the requirements...

Страница 14: ...icrosoft Windows and that you are familiar with basic file management within Windows Required Equipment PC 9 pin null modem cable with female connectors on both ends The 9 pin null modem cable can be made from a standard female female 9 pin serial cable and a correctly gendered null modem adapter or it can be assembled from the following parts Item Quantity Description Part Number 1 1 9 pin Serial...

Страница 15: ...signal generator error messages list on the documentation CD for information about each error message After resolving all problems causing errors press Clear Error Queue s If the ERR annunciator is off the signal generator functionality check has passed Electrostatic discharge ESD can damage or destroy electronic components All work on electronic assemblies should be performed at a static safe wor...

Страница 16: ...wing the download instructions see Helpful Notes for Using this Procedure on page 17 10 Once the firmware download is successful disconnect the RS 232 cable 11 The files and directories that were created in the earlier steps can now be removed from your computer For firmware revisions B 02 42 and above the download procedure has changed It is no longer necessary to synchronize the powering up of t...

Страница 17: ... are prompted with Be sure the instrument RS 232 settings under the front panel keys Utility HP IB RS 232 are set as follows turn the signal generator on and verify the RS 232 settings as shown then turn the signal generator off When you are prompted with Do you have Option UND UN5 and or H97 Y N Return answering Yes will download the ESG firmware and Option UND UN5 and or H97 firmware Answering N...

Страница 18: ...This information is subject to change without notice Keysight Technologies Inc 1999 2015 Edition 1 Month 2015 E4400 90549 www keysight com ...

Отзывы: