background image

Memory Erase Processes: Product model, versions, calibration data, calibration count contains no application-specific information. 
Memory-stored data, IO configuration, register configuration and states, calibration secure state, calibration secure code, calibration 
message are user-saved data. See below for memory erase processes. 

 

Memory-stored data 
NOTE: Make sure the instrument is not in the calibration mode. 
On the front panel:  

1.

 

Press STORE > NAME STATE > 1: XXXXXX > 1:STATE1 > DONE/NO CHANGE. 

2.

 

Press STORE > NAME STATE > 2: XXXXXX > 2:STATE2 > DONE/NO CHANGE. 

3.

 

Press STORE > NAME STATE > 3: XXXXXX > 3:STATE3 > DONE/NO CHANGE. 

4.

 

Press STORE > NAME STATE > 4: XXXXXX > 4:STATE4 > DONE/NO CHANGE. 

5.

 

Press STORE > NAME STATE > 5: XXXXXX > 5:STATE5 > DONE/NO CHANGE. 

6.

 

Power cycle the unit. 

7.

 

Press STORE > STORE STATE > 1: STATE1 > DONE. 

8.

 

Press STORE > STORE STATE > 2: STATE2 > DONE. 

9.

 

Press STORE > STORE STATE > 3: STATE3 > DONE. 

10.

 

Press STORE > STORE STATE > 4: STATE4 > DONE. 

11.

 

Press STORE > STORE STATE > 5: STATE5 > DONE. 

 

From the remote interface, send the following command in sequence: 

1.

 

MEMory:STATe:NAME 1, "STATE1" 

2.

 

MEMory:STATe:NAME 2, "STATE2

 

3.

 

MEMory:STATe:NAME 3, "STATE3" 

4.

 

MEMory:STATe:NAME 4, "STATE4" 

5.

 

MEMory:STATe:NAME 5, "STATE5" 

6.

 

*RST 

7.

 

*SAV 1 

8.

 

*SAV 2  

9.

 

*SAV 3 

10.

 

*SAV 4 

11.

 

*SAV 5 

 

IO configuration 
On the front panel: 

1.

 

Press I/O Config > RS-232 > 9600 BAUD > NONE 8 BITS > CHANGE SAVED/NO CHANGE. 

2.

 

Press I/O Config > GPIB / 488 > ADDR 05 > SAVED/NO CHANGE. 

 

Register configuration and state 
From the remote interface, send *PSC 1, then power cycle the unit. 

 

Calibration secure state, calibration secure code and calibration message 
From the remote interface, send the following command in sequence:  

-

 

CALibration:SECure:STATe OFF, <code> 
NOTE: <code> refers to 

0036

40” for model E3640

A, "003641" for model E3641A, "003642" for model E3642A, "003643" 

for model E3643A,

 

"003644" for model E3644A, and "003645" for model E3645A. Applicable only if user has not changed 

the model number from the default value. 

-

 

CALibration:SECure:CODE <new code> 
NOTE: <new_code> must be "003640" for model E3640A, "003641" for model E3641A, "003642" for model E3642A, 
"003643" for model E3643A, "003644" for model E3644A, and "003645" for model E3645A. 

-

 

CALibration:STRing "" 
NOTE: This resets the calibration message to an empty string. 

-

 

CALibration:SECure:STATe ON, <code> 
NOTE: <code> must be "003640" for model E3640A, "003641" for model E3641A, "003642" for model E3642A, "003643" 
for model E3643A, "003644" for model E3644A, and "003645" for model E3645A. 

 

 
 

Содержание E3640A

Страница 1: ...on code storage User Modifiable Y N N Volatile Y N N Memory Erase Processes Not applicable contains no application specific information Memory Type RAM Digital Board Memory Size 256 kbit Memory Function Digital microcontroller temporary execution data User Modifiable Y N N Volatile Y N Y Memory Erase Processes On power cycle Memory Type ROM Digital Board Memory Size 4 kbit Memory Function Product ...

Страница 2: ... Type RAM Front panel Memory Size 4 kByte Memory Function Front panel microcontroller temporary execution data User Modifiable Y N N Volatile Y N Y Memory Erase Processes On power cycle Memory Type ROM IO controller Memory Size 12 kByte Memory Function IO controller microcontroller execution code storage User Modifiable Y N N Volatile Y N N Memory Erase Processes Not applicable contains no applica...

Страница 3: ...NAME 3 STATE3 4 MEMory STATe NAME 4 STATE4 5 MEMory STATe NAME 5 STATE5 6 RST 7 SAV 1 8 SAV 2 9 SAV 3 10 SAV 4 11 SAV 5 IO configuration On the front panel 1 Press I O Config RS 232 9600 BAUD NONE 8 BITS CHANGE SAVED NO CHANGE 2 Press I O Config GPIB 488 ADDR 05 SAVED NO CHANGE Register configuration and state From the remote interface send PSC 1 then power cycle the unit Calibration secure state ...

Страница 4: ... process could be available with engineering resources Sanitization process is not possible i e sanitize the product by destruction This information is subject to change without notice Keysight Technologies 2018 Edition 1 October 2018 Printed in Malaysia E3640 90002 www keysight com ...

Отзывы: