Initial Operation
2
Keysight E3633A and E3634A User’s Guide
41
Power-on Checkout
The power-on test includes an automatic self-test that checks the internal
microprocessors and allows the user visually to check the display. You will
observe the following sequence on the display after pressing the front panel
power switch to on.
1 The front-panel display will light up briefly while the instrument performs
its power-on self-test.
To review the power-on display with all annunciators turned on, hold down
key as you turn on the power supply.
2 The GPIB address or RS-232 is also displayed for about one second.
The GPIB address is set to “5” when the power supply is shipped from the
factory for remote interface configuration. If this is not the first time the power
supply is turned on, a different interface (RS-232) or a different GPIB address
may appear.
See
“Remote Interface Configuration”
if you need to
change the remote interface configuration.
3 The “8V”
[1]
or “25V”
[2]
, “OVP”, “OCP” and “OFF” annunciators are on. All
others are off.
The power supply will go into the power-on / reset state; the output is disabled
(the
OFF
annunciator turns on); the 8V/20A
or 25V/7A
range is selected
(the
8V
or
25V
annunciator turns on); and the knob is selected for voltage
control. Notice that the
OVP
and
OCP
annunciator also turn on.
ADDR (or RS-232)
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Содержание E3633A
Страница 1: ...Keysight E3633A and E3634A DC Power Supplies User s Guide ...
Страница 10: ...10 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...20 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 133: ...Remote Interface Reference 4 Keysight E3633A and E3634A User s Guide 133 SCPI Status System ...
Страница 166: ...5 Error Messages 166 Keysight E3633A and E3634A User s Guide 631 System DAC test failed 632 Hardware test failed ...
Страница 170: ...5 Error Messages 170 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 204: ...8 Specifications 204 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...