30
Keysight E1412A 6 1/2 Digit Multimeter Service Guide
Verification Tests
Performance Verification
AC Voltage Gain Verification
Connect the AC Standard across the multimeter HI and LO input terminals as
shown in Figure 4-5.
1
Set up the multimeter as follows:
Reset the multimeter
*RST
Configure for AC voltage, 100mV
CONF:VOLT:AC 0.1
Select Slow AC filter
DET:BAND 3
2
Set the AC Standard output to 100 mV at 1 kHz.
Table 4-4
AC Voltage Gain Verification Tests
INPUT
INPUT
FREQUENCY
Keysight
E1412A Range
QUICK
VERIFICATION
Error from nominal
24 hour
90 day
1 year
10 mV
1 kHz
100 mV
± 34 μV
± 45 μV
± 46 μV
100 mV
1 kHz
100 mV
± 70 μV
± 90 μV
± 100 μV
100 mV
50 kHz
100 mV
Q
± 150 μV
± 160 μV
± 170 μV
1 V
1 kHz
1 V
± 600 μV
± 800 μV
± 900 μV
1 V
50 kHz
1 V
± 1.4 mV
± 1.6 mV
± 1.7 mV
10 V
10 Hz
10 V
± 37 mV
± 38 mV
± 38 mV
10 V
1 kHz
10 V
± 6 mV
± 8 mV
± 9 mV
10 V
50 kHz
10 V
Q
± 14 mV
± 16 mV
± 17 mV
100 V
1 kHz
100 V
Q
± 60 mV
± 80 mV
± 90 mV
100 V
50 kHz
100 V
± 140 mV
± 160 mV
± 170 mV
300 V
1 kHz
300 V
± 300 mV
± 420 mV
± 450 mV
300 V
1
50 kHz
300 V
± 660 mV
± 780 mV
± 810 mV
1 Some calibrators may have difficulty driving the multimeter and cable load at this V-Hz output. Use short, low-capacitance cable
to reduce calibration loading. Verification can be performed at >195 Vrms. New test limits can be computed from the accuracy
specification shown in chapter 1 for the actual test conditions used.
Q = Quick performance verification test points.
The 50 kHz ac voltage test points may fail performance
verification if the internal shields have been removed and
reinstalled. You must recalibrate the ac voltage function any time
the multimeter’s internal shields have been removed and
reinstalled. Consult the “AC Voltage Gain Adjustment” to
recalibrate the ac voltage function.
Содержание E1412A
Страница 1: ...Keysight E1412A 6 1 2 Digit Multimeter Service Manual 75000 Series C ...
Страница 2: ......
Страница 6: ...vi ...
Страница 12: ...xii Keysight E1412A 6 1 2 Digit Multimeter Service Guide ...
Страница 22: ...16 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Operating Instructions Warnings and Cautions ...
Страница 74: ...68 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Adjustments CALibration ...
Страница 80: ...74 E1412A 6 1 2 Digit Multimeter Service Guide Backdating ...
Страница 117: ...110 E1412A 6 1 2 Digit Multimeter Service Guide Keysight E1412A Multimeter Error Messages Execution Errors ...
Страница 118: ......