208
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
12
Trigger Mode/Coupling
To select the Auto or Normal trigger mode
When the oscilloscope is running, the trigger mode tells the oscilloscope what to
do when triggers are not occurring.
In the
Auto
trigger mode (the default setting), if the specified trigger conditions are
not found, triggers are forced and acquisitions are made so that signal activity is
displayed on the oscilloscope.
In the
Normal
trigger mode, triggers and acquisitions only occur when the
specified trigger conditions are found.
To select the trigger mode:
1
Press the
[Mode/Coupling]
key.
2
In the Trigger Mode and Coupling Menu, press the
Mode
softkey; then select
either
Auto
or
Normal
.
You can also make this selection using the touchscreen. See
Trigger Menu, Change the Trigger Mode, and Open the Trigger Level
See the following
“When to Use Auto Trigger Mode"
on page 209 descriptions.
You can also configure the
[Quick Action]
key to toggle between the Auto and
Normal trigger modes. See
“Configuring the [Quick Action] Key"
Triggering and the
Pre- and
Post-Trigger
Buffers
After the oscilloscope starts running (after pressing
[Run]
or
[Single]
or changing
the trigger condition), the oscilloscope first fills the pre-trigger buffer. Then, after
the pre-trigger buffer is filled, the oscilloscope starts searching for a trigger, and
sampled data continues to flow data through the pre-trigger buffer in a first-in
first-out (FIFO) manner.
When a trigger is found, the pre-trigger buffer contains the events that occurred
just before the trigger. Then, the oscilloscope fills the post-trigger buffer and
displays the acquisition memory. If the acquisition was initiated by
[Run/Stop]
, the
process repeats. If the acquisition was initiated by pressing
[Single]
, the acquisition
stops (and you can Pan and Zoom the waveform).
In either Auto or Normal trigger mode, a trigger may be missed if the event occurs
while the pre-trigger buffer is being filled. This may be more likely, for example,
when the horizontal scale knob is set to a slow time/div setting, such as
500 ms/div.
Содержание DSOX3012T
Страница 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Страница 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Страница 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Страница 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Страница 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Страница 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Страница 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Страница 294: ...294 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 17 Digital Voltmeter and Counter...
Страница 300: ...300 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 18 Frequency Response Analysis...
Страница 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Страница 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Страница 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Страница 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...
Страница 416: ...416 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 25 CAN LIN Triggering and Serial Decode...
Страница 426: ...426 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 26 CXPI Triggering and Serial Decode...
Страница 436: ...436 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 27 FlexRay Triggering and Serial Decode...
Страница 480: ...480 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 30 Manchester NRZ Triggering and Serial Decode...
Страница 510: ...510 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 32 SENT Triggering and Serial Decode...
Страница 520: ...520 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 33 UART RS232 Triggering and Serial Decode...