Keysight CX3300 User’s Guide, Edition 4
75
Using the CX3300
Waveform Analytics
Performing Waveform Analytics by Waveform Recording
You can perform Waveform Analytics during the waveform recording as shown
below.
1. Open the Data Logger dialog box (Data Logger > Recording > Trigger) and set
the slide switch of Trigger for Triggered Segmentation to ON.
2. Set the segment length mode to “Single”, “Dual”, or “Trigger to Trigger”.
3. Define the trigger settings.
The “Import from Hardware Settings” button allows you to use the trigger
settings of the Scope mode.
The “Generate from Displayed Waveform” button allows you to use the
trigger settings based on the waveform displayed.
4. Open the Data Logger dialog box (Data Logger > Recording > Similarity) and
set the Similarity Threshold. Available values are 0% to 99.9999%.
5. Click the “Start Recording” button (Data Logger > Recording).
When the waveform recording is completed, the Quick clustering is performed.
Then the waveform database will be automatically loaded, and the Waveform
Analytics tool palette will appear automatically. This tool palette shows the
Quick clustering results.
Performing Waveform Analytics by Retriggering
You can perform Waveform Analytics for the existing waveform database by
using Retriggering without performing the measurement.
1. Load the existing waveform database.
2. Open the Data Logger dialog box (Data Logger > Retriggering) and set the
slide switch of Trigger for Triggered Segmentation to ON.
3. Define the trigger settings.
4. Set the Similarity Threshold. Available values are 0% to 99.9999%.
5. Click the “Start Retriggering” button. The Retriggering dialog box will open,
and you can change the record name if necessary. The record name must
contain only ASCII characters. Then, click the “Yes” button to start
Retriggering.