16
Keysight CX3300 User’s Guide, Edition 4
Setting Up the CX3300
Installation
Table 1-1
Environmental Characteristics
Environment
Indoor use only
Ambient temperature
Operating: 0 ˚C to +40 ˚C
Storage: −20 ˚C to +60 ˚C
Humidity
Operating: up to 80% relative humidity, non-condensing, at +40 ˚C
Storage: up to 90% relative humidity, non-condensing, at +60 ˚C
Altitude
Operating: up to 2,000 m (6,561 ft.)
Storage: up to 4,600 m (15,092 ft.)
Weight
11 kg
Dimensions
425.6 mm (W) × 266.1 mm (H) × 196.7 mm (D), with feet retracted
Safety
IEC 61010-1
EMC (Electromagnetic Compatibility)
IEC 61326-1
Measurement category
Measurement category 1
Power requirements
100—240 V (±10%), 50/60 Hz
250 VA, maximum power dissipated