132
Keysight B2980B User’s Guide, Edition 1
Front Panel Reference
Hard Keys and Rotary Knob
Hard Keys and Rotary Knob
Standby switch
Turns the instrument on or off. During the power is on, the LED below the switch
turns green. For the B2983B/B2987B, the LED turns orange during the switch is
off and the battery charging is in progress.
Single
Starts a single (one shot) measurement or initiates trigger system. Also triggers
the staircase sweep output or square waveform output of B2985B/B2987B. If a
repeat measurement is in progress, it stops it and starts a single measurement.
The measurement result is displayed on the Meter view, Graph view, Histogram
view, or Roll view.
Once a single measurement starts, the data buffer is cleared and stores the last
single measurement result (maximum 100000 data). It can be listed and displayed
on the Measure Result dialog box.
If measurement is not performed properly, check the trigger setting. The trigger
type must be set to AUTO, or the trigger count (Count) must be set properly. See
“Trigger Setup Sub-panel” on page 157
Run/Stop
Starts a repeat (continuous) measurement. If a repeat measurement is in progress,
this stops it. The measurement result is displayed on the Meter view, Histogram
view, or Roll view.
Once a repeat measurement starts, the data buffer is cleared and stores the last
repeat measurement result (maximum 1000 data). It can be listed and displayed
on the Measure Result dialog box.
The repeat measurement is performed with the following trigger setup. Settings of
are ignored.
•
Acquire trigger (measurement trigger): Initiate
•
ARM acquire count: Infinite
•
ARM acquire source: AUTO (automatic internal)
•
TRIGger acquire count (Measure Count): 100
•
TRIGger acquire source (Measure Trigger): AUTO (automatic internal)
Содержание B2980B Series
Страница 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Страница 16: ......
Страница 24: ......
Страница 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Страница 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Страница 130: ...130 Keysight B2980B User s Guide Edition 1 How to Perform Measurement Measurement Considerations...
Страница 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Страница 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...