07 | Keysight | Using Source/Measure Unit as an Ammeter – Demo Guide
2-3. Perform the measurement
1) Press for the channel 1 to switch on its output terminal.
2) Press to perform a measurement repeatedly. Now you can see the
measurement result on the GUI of the B2902/12A as below.
3. Enable the DUT, the current source circuit
You can also use your own power supply instead of the channel 2 of the B2902A/12A
3-1. Change View mode to Ch2 Single View.
1) Press repeatedly until the
Channel 2
Single View
is displayed.
3-2. Configure the condition of the power supply (Channel 2)
1) Press to edit the
Source function
, and then select to set the
Source
function
to
Voltage source.
2) Press to edit the
Source value
, and then enter
1.1 V
to set the
Source
value
to
1.1 V
for example.
3) Press to edit the
Limit value
, and then enter
100 mA
to set the
Limit value
to
100 mA
for example.