Environmental Specifications
The N1923/4A low PIM coaxial multiport switches are designed to fully comply with Keysight’s
product operating environmental specifications.
Table 3-7
Environmental specifications
Parameter
Specification
Temperature
– Operating
– Storage
– Cycling
–25
°
C to 75
°
C
–55
°
C to 85
°
C
–50
°
C to 150
°
C, 10 cycles
Humidity
– Operating
– Storage
– Condensation
40
°
C/95% RH, 5 days
65
°
C/90% RH, 24 hours
40
°
C/95% RH
Shock
– Non-operating:
– Half-sine
– Transportation
– Operating
500 G at 0.5 ms, 3 drops/direction
50 G Vibration: 8 m/s
±
10%
50 G at 6 ms, 6 directions
Vibration
– Operating
– Survival
– Random
7 G rms, 5 to 2000 Hz at 0.25 in p-p
20 G rms, 20 to 2000 Hz at 0.06 in p-p, 4 min/cycle, 4 cycles/axis
7 G rms, 50 to 2000 Hz, 15 min/axis
ESD immunity
– Direct discharge
– Air discharge
6 kV (to outer conductor)
15 kV (to outer conductor)
RFI
Radiated emission per CISPR 11
Magnetic field
– Operating emission
– Operating immunity
AC magnetic emission (1.88 G rms)
DC magnetic emission (5 G)
30 A/M rms at 47 Hz, 50 Hz, 60 Hz, and 189 Hz
150 A/M rms at 47 Hz and 189 Hz
Содержание 8710 Series
Страница 1: ...Keysight 8710x Low PIM Coxial Multiport Switches Operating and Service Manual ...
Страница 4: ...4 Keysight 8710x Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 6: ...6 Keysight 8710x Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 8: ...8 Keysight 8710x Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 10: ...10 Keysight 8710x Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...1 Introduction 14 Keysight 8710x Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 23: ...Specifications 3 Keysight 8710x Operating and Service Manual 23 Figure 3 2 Power derating factor vs VSWR ...