3
Performance Tests
36
Keysight 86205A Operating and Service Manual
6
When prompted, connect the calibration devices to port 1 of the bridge as shown in
(measure the load last).
Figure 3-8
Equipment setup for S22 1-port measurement calibration
7
Load the frequency response error term into memory when the calibration is done, by pressing:
SYSTEM
SERVICE MENU, TESTS
34, x1, EXECUTE TEST
DISPLAY, DATA/MEM
CAL, CORRECTION OFF
The resulting display is the uncorrected load data (the last measurement taken during the
calibration), normalized to the frequency response of the system.
8
Find the worst-case directivity value, by pressing:
MKR FCTN
MARKER SEARCH MAX
NOTE
When measuring the open and short devices, select OPEN (f) and SHORT (f) on the analyzer. The
selection made must correspond to the test port
not
the device being measured.
Measure the open and short circuit devices; then measure the load.
KEYSIGHT 8753C
NETWORK
ANALYZER
KEYSIGHT 86025A
DIRECTIONAL
BRIDGE
Содержание 86205A
Страница 1: ...Keysight 86205A Directional Bridge Operating and Service Manual...
Страница 4: ...4 Keysight 86205A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 6: ...6 Keysight 86205A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 8: ...8 Keysight 86205A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 10: ...10 Keysight 86205A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 38: ...3 Performance Tests 38 Keysight 86205A Operating and Service Manual Figure 3 9 86205A test record 1 of 3...
Страница 40: ...3 Performance Tests 40 Keysight 86205A Operating and Service Manual Figure 3 11 86205A test record 3 of 3...
Страница 44: ...4 Maintenance 44 Keysight 86205A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...