Scanner Maintenance and Calibration
5
Keysight 5500 SPM User’s Guide
5-14
If the measured dimension does not match the actual, then adjust the X
Sensitivity term according to the following equation:
After adjustment the measured size should be within 5 % of the actual
size.
Y Calibration
The methods used for calibrating the scanner’s Y axis are similar to
those used for the X dimension. The exception is that the scan range will
be set to ½ of the scanner’s full range. The Y axis is generally used as
the slow scanning axis so the range will be reduced as a time
consideration.
Only one Topography image is required for Y calibration. The other
image can be assigned to display flattened Topography data.
Y Non-Linearity
Obtain an image of the calibration target. Using the Cross Section tool,
set markers at the uppermost and bottommost feature along a vertical
cross section (
Figure 8-11
Cross-section in Y direction showing non-linearity
NewSensitivity
CurrentSensitivit y Kno
wnSize
MeasuredSize
--------------------------------------------------------------------------------------
=
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...