Formats and Data Flow
7
Keysight 53210A User’s Guide
195
Front Panel Display
Counter measurements taken from the
front panel
or from an IO interface appear
in the display as indicated in
for an overview of display
features.
Reading Memory
In addition to being displayed, all counter readings are stored in the counter’s
volatile
reading memory
). Memory capacity is 1,000,000 readings.
Reading memory has the following attributes:
1
reading memory is cleared after a function change, during execution of the
READ?
,
INITiate:IMMediate
, or
MEASure?
command, after cycling power, or
following an instrument reset (
*RST
) or preset (
SYSTem:PRESet
).
Table 7-2
Commands Affecting Data Flow
Command
Locations
Accessed
Readings
Transferred
Reading Memory
Cleared
Allowed during
Measurement
Read?
Front Panel
Reading Memory
Output Buffer
All
No
No
INITiate:IMMediate
Front Panel
Reading Memory
None
Yes
No
FETCh?
Reading Memory
Output Buffer
All
No
No
R?
Reading Memory
Output Buffer
All or
specified count
Yes
Yes
DATA:REMove?
Reading Memory
Output Buffer
Specified Count
Yes
Yes
DATA:LAST?
Reading Memory
Output Buffer
1
No
Yes
MMEMORY:STORe
:DATA
Reading Memory
Internal Flash
External USB
All
No
No
Содержание 53210A
Страница 1: ...Keysight 53210A 350 MHz RF Counter User s Guide ...
Страница 12: ...12 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...1 Preparation for Use 24 Keysight 53210A User s Guide Figure 1 1 The 53210A Universal Frequency Counter Timer ...
Страница 48: ...1 Preparation for Use 48 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 122: ...4 53210A Input Signal Conditioning 122 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 222: ...8 Instrument Status 222 Keysight 53210A User s Guide Figure 8 1 The 53210A Status System Instrument Locked Global Error ...
Страница 228: ...8 Instrument Status 228 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 240: ...A 53210A Error Messages 240 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 246: ...246 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...