COMMAND
:MEMory:CLEAR[:NAME] . . . <name>
Resets the contents of the named power calibration table to the default
configuration.
COMMENTS
– Valid table names:
8481A
,
8481D
,
8482A
,
8485A
,
8487A
,
CUST1
,
CUST2
,
CUST3
– The data currently stored in non-volatile memory for the named calibration
table is discarded.
– This is an event. There is no query form of this command.
RELATED
FRONT-PANEL
KEYS
Shift + Menu
(HEAD > OFF
,
8481A
,
8481D
,
8482A
,
8485A
,
8487A
,
CUST1
,
CUST2
,
CUST3)
COMMAND
:MEMory:NSTates?
Queries the Number of available
*SAV/*RCL
STates
in the instrument.
QUERY
RESPONSE
– Numeric data transferred as ASCII bytes in
<NR1>
format.
– The value returned is 9.
– The response value is one greater than the maximum which can be sent as a
parameter to the
*SAV
and
*RCL
commands.
COMMENTS
Query only.
RELATED
FRONT-PANEL
KEYS
None