Programming
3
Keysight 53147A/148A/149A Operating and Programming Guide
213
RELATED
FRONT-PANEL
KEYS
Shift + Cal Factor
COMMAND
:CALibration:ZERO:AUTO ONCE
This command causes the power meter to perform its zeroing routine.
This adjusts the power meter for a zero power reading with no power supplied to
the power sensor.
COMMENTS
– The command assumes that the power sensor is not connected to a power
source.
– Pressing the STORE key sets
CAL:RCF
to the frequency displayed on the
front-panel display.
– After
*RST
, automatic zeroing is disabled.
RELATED
FRONT-PANEL
KEYS
Shift + Zero; Shift + Store
COMMAND
:CALibration:ZERO:AUTO?
QUERY
RESPONSE
The query always returns a value of 0.
– If zeroing was not successful, the error
–340, “Data Questionable; ZERO
ERROR”
occurs.
– If this command is set to
1
| ON the error
–224, “Illegal parameter value”
occurs.