2
Operating Your Instrument
102
Keysight 53147A/148A/149A Operating and Programming Guide
lists the factory-set values in all of the data points for all five of the
pre-recorded power head calibration tables.
Table 2-2
Calibration Table Data-Point Values
Power Head Model Number
Frequency
Calibration Factor
8481A
RCF
100.0
50 MHz
100.0
100 MHz
99.8
2 GHz
99.0
3 GHz
98.6
4 GHz
98.0
5 GHz
97.7
6 GHz
97.4
7 GHz
97.1
8 GHz
96.6
9 GHz
96.2
10 GHz
95.4
11 GHz
94.9
12.4 GHz
94.3
13 GHz
94.3
14 GHz
93.2
15 GHz
93.0
16 GHz
93.0
17 GHz
92.7
18 GHz
91.8