64
4000 X-Series Oscilloscopes Advanced Training Guide
3
Advanced Triggering, Search & Navigation, and Segmented Acquisition Labs
Your scope’s display should look similar to
showing a positive runt pulse
that is approximately 100 ns wide near center-screen. In addition to triggering on
runt pulse conditions, the Keysight 4000 X-Series oscilloscopes can also perform
Search & Navigation
to find multiple runt pulse conditions, regardless of the specific
trigger setup condition. Let’s now capture a longer stream of pulses and perform
an automatic search operation to find all runts.
15
Set the scope’s timebase to
100.0 µs/div
.
16
Press
[Run/Stop]
to stop repetitive acquisitions.
17
Press the
[Search]
front panel key.
18
Tap the
Search Edge
softkey; double-tap on
Runt
.
19
Tap the
Settings
softkey to observe the default runt search conditions.
Figure 38
Triggering on a positive runt pulse that is < 200 ns wide.